STARTER 3100C Bench Conductivity Meter EN-5
2 INSTALLATION
Carefully unpack the meter.
2.1 Package
contents
The ST3100C-B package has the following items
In addition to ST3100C-B content, the ST3100C-F package also includes the following:
The best measurement range for STCON3 4-ring (4-pole) conductivity electrode is
70
μ
S/cm
-200
mS/cm. In this range the measurement accuracy can reach 0.5%.
If you use STCON3 to measure pure water in which conductivity value is below 70
μ
S/cm
; the
measurement will be less accurate (approximately 1% to 5%).
Does not use STCON3 to measure cond. value less than 2
μ
S/cm
, the accuracy is not good.
Accessories:
Model
Description
P/N
STCON3
4-ring cond. electrode
(
70
μ
S/cm
-200mS/cm, 0.5% accuracy; )
( 2-70
μ
S/cm
, 1-5% accuracy)
83033972
STCON7
Pure water cond. electrode
(0.02
μ
S/cm
-200
μ
S /cm
; 0.02
μ
S
/cm accuracy)
30080693
Standard Conduct 10
μ
S/cm 250ml
30100441
Standard Conduct 84
μ
S/cm 250ml
30100442
Standard Conduct 1413
μ
S/cm 250ml
30100443
Standard Conduct12.88mS/cm 250ml
30100444
Electrode holder stand alone (3100/3100C)
30058733
In use cover
30058734
ST3100C-B
Units
STARTER 3100C meter
1
Stand alone electrode holder
1
1413
μ
S/cm standard solution
1 bottle (about 20ml)
12.88 mS/cm standard solution
1 bottle (about 20ml)
In use cover
1
STCON3 conductivity electrode
1
www.
GlobalTestSupply
.com
Find Quality Products Online at: