EN-16 STARTER 3100C Bench Conductivity Meter
6 TECHNICAL
DATA
6.1 Specifications
Ambient conditions
x
Indoor use only
x
Altitude: Up to 2000 m
x
Specified Temperature range: 5°C to 40°C
x
Humidity: maximum relative humidity 80 % for temperatures up to 30°C decreasing linearly to
50% relative humidity at 40°C
x
Mains supply voltage fluctuations: up to ±10% of the nominal voltage
x
Installation category II
x
Pollution degree: 2
x
Operability is assured at ambient temperatures between 5°C to 40°C
Model
ST3100C
Measuring range
0.0
μ
S/cm…199.9 mS/cm
0.1 mg/l…199.9 g/l (TDS)
0.00 ... 100.0 psu (Salinity)
0 °C…100 °
C
Resolution
Automatic range
0.1 °C
Error limits/Accuracy
± 0.5 % of the measured value
± 0.3 °C
Calibration
1 point
4 predefined standards
Memory
99 measurements
The last calibration data
Power supply
AC Adapter Input: 100-240V ~ X.XA 50/60 Hz
AC Adapter Output: 12V
¯
X.XA
Size/weight
220 W x 175 D x 78 H mm / 0.75 kg
Display
White backlit Liquid crystal
Input
Mini-Din
Temperature-compensation
Linear: 0.00 %/°C…10.00 %/°
C
Reference temperature: 20 & 25 °C
Housing
ABS
www.