
STARTER 3100C Bench Conductivity Meter EN-9
Use
or
buttons to adjust the temperature correction value. (0.00 to 10.00 %/°C).
Press
to confirm your setting. Continue with reference temperature setting by following
or press
to go back to the measurement screen.
Note:
If you set the temperature correction coefficient value as 0 %/°C , that means
NO Temperature
Compensation
for the conductivity measurement. The meter displays the real conductivity value
at current temperature.
3.5
Set TDS factor
After selecting the reference temperature, the current TDS factor blinks.
Use
or
buttons to adjust the TDS factor value. Press
to confirm your
setting or press
to go back to the measurement screen.
www.
GlobalTestSupply
.com
Find Quality Products Online at: