STARTER 3100C Bench Conductivity Meter                               EN-9

 

 

 

 

Use 

 or 

 

buttons to adjust the temperature correction value. (0.00 to 10.00 %/°C). 

Press 

  to confirm your setting. Continue with reference temperature setting by following 

or press 

  to go back to the measurement screen.  

 

Note: 

If you set the temperature correction coefficient value as 0 %/°C , that means 

NO Temperature 

Compensation

 for the conductivity measurement. The meter displays the real conductivity value 

at current temperature. 
 

3.5 

Set TDS factor 

After selecting the reference temperature, the current TDS factor blinks.  

Use 

 or 

 

buttons to adjust the TDS factor value. Press 

   to  confirm  your 

setting or press 

  to go back to the measurement screen. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Summary of Contents for ST3100C-B

Page 1: ...e instrucciones del Conductiv metro de mesa STARTER 3100C STARTER 3100C Appareil de mesure de la conductivit de laboratoire Manuel d instructions STARTER 3100C Medidor de Condutividade de Bancada Manu...

Page 2: ...5 Set TDS factor 9 4 STARTER 3100C OPERATION 10 4 1 Calibration 10 4 1 1 Selecting a standard 10 4 1 2 Performing a calibration 10 4 2 Sample measurement 11 4 3 TDS and Salinity measurement 11 4 4 Usi...

Page 3: ...Signal Warnings and Symbols Safety notes are marked with signal words and warning symbols These show safety issues and warnings Ignoring the safety notes may lead to personal injury damage to the ins...

Page 4: ...al obstacle or tripping hazard x Use the equipment only in dry locations x Dry off any liquid spills immediately The instrument is not watertight x When using chemicals and solvents comply with the in...

Page 5: ...unning 3 Calibration icon calibration in progress 4 Setup icon instrument is in the setup mode 5 TDS factor in the setup mode 6 Conductivity TDS Salinity Cell Constant 7 Standard Error index 8 Tempera...

Page 6: ...t Meter turn on Return to measurement screen Meter turn off Store current reading to memory Increase value during setting Scroll up through the memory Recall stored data Print current memory data Swit...

Page 7: ...than 2 S cm the accuracy is not good Accessories Model Description P N STCON3 4 ring cond electrode 70 S cm 200mS cm 0 5 accuracy 2 70 S cm 1 5 accuracy 83033972 STCON7 Pure water cond electrode 0 02...

Page 8: ...u can move up and down When the electrode is installed on the arm the storage bottle of pH electrode fits into the base hole 2 3 Installing the power adapter Insert the right adapter clip into the pow...

Page 9: ...RTER 3100C is the meter stand which can be used to adjust the view angle of the display in case of working on high lab table 2 6 Attached quick guide Another unique design of the STARTER 3100C is the...

Page 10: ...e setting Use or buttons to adjust the temperature value Press to confirm your setting Continue with calibration standard setting by following or press to go back to the measurement screen 3 3 Set cal...

Page 11: ...ure correction coefficient value as 0 C that means NO Temperature Compensation for the conductivity measurement The meter displays the real conductivity value at current temperature 3 5 Set TDS factor...

Page 12: ...and value Then the current standard blinks Use or buttons to select the standards press to confirm Press to leave the setup mode The 4 predefined standards are 10 S cm 84 S cm 1413 S cm 12 88 mS cm T...

Page 13: ...rify your cell constant with a standard solution before measurement and recalibrate if necessary Always use fresh standards For STCON3 the normal cell constant range is 1 50 2 00 e g 1 71 cm if the ce...

Page 14: ...ndicates which result is being displayed Press to exit 4 4 3 Clearing the memory Pressing or to scroll through the stored results until MRCL appears Press CLr blinks Now we can two choices Press to co...

Page 15: ...at for the print out following a Salinity measurement is End Point Value Temp ATC MTC Manual EP 0 08psu 25 0 C MTC The print out in case of an error message is End Point Value Temp ATC MTC Error4 The...

Page 16: ...eep 3 times to alert 5 2 Meter maintenance Never unscrew the two halves of the housing The STARTER 3100C does not require any maintenance To clean use a damp cloth The housing is made of acrylonitrile...

Page 17: ...Recover Factory Setup When the meter is off press and hold together for 3 seconds the screen displays and blinks means Reset Then we have 2 choice Press to reset factory settings MTC slope and offset...

Page 18: ...atures between 5 C to 40 C Model ST3100C Measuring range 0 0 S cm 199 9 mS cm 0 1 mg l 199 9 g l TDS 0 00 100 0 psu Salinity 0 C 100 C Resolution Automatic range 0 1 C Error limits Accuracy 0 5 of the...

Page 19: ...ent has been tested and found to comply with the limits for a Class A digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful int...

Page 20: ...Cl 10 2 14 H2SO4 10 1 28 HF 1 5 7 20 coefficients of conductivity standards for a calculation to reference temperature of 25 C Standard Measurement temp 15 C Measurement temp 20 C Measurement temp 30...

Page 21: ...use exposed to radioactive or corrosive materials has foreign material penetrating to the inside of the product or as a result of service or modification by other than Ohaus In lieu of a properly retu...

Reviews: