3.4 Weighing

 

Mode

 

This mode is the factory default setting.  

1. If needed, press and hold 

Mode

 until [

wWeIGH

] (Weigh) is displayed. 

2. If required, place an empty container on the pan and press 

Tare

3. Add sample to the pan or container. The display shows the weight of the sample. 
 

3.5 Counting Mode 

 

This mode counts large numbers of items based on the weight of a reference count. 

1.   Place an empty container on the pan and press 

Tare

.

 

2.   Press and hold 

Mode

 until [

Count

] (Count) is displayed. [

CLr.PwW

] (Clear Average Piece 

  Weight, APW) will then display.  

    If no APW exists, the balance will display [

Put.10

], proceed to step 5. 

3.   Press 

No

 to use the stored APW. Proceed to step 7. 

4.  Press 

Yes

 to establish an APW.  

5.  The balance will then display the stored sample size, i.e. [

Put 10

]. Press 

No

 or

 Back

 to 

toggle the choices (5,   10, 20, 50 or 100). 

6.   Put the indicated number of pieces on the pan then press 

Yes 

to calculate the APW. The 

  display shows  the piece count. 

7.  Add additional pieces until the desired count is reached.  
8.  To clear the stored APW press and hold 

Mode

 until [

Count

] is displayed. Press 

Yes

 when      

       [

Clr.pwW

] is displayed.  

   Note: Press 

Function

 to view the current APW. 

 
3.6 Percent Mode 

This mode measures the weight of a sample as a percentage of a reference weight. 

1.   If required place an empty container on the pan and press 

Tare

2.   Press and hold 

Mode

 until [

Percnt

] is displayed. [

Clr.ref

]

 

(clear reference) will then  

  display. If no reference weight exists, the balance will display [

Put.ref

], proceed to step 5. 

3.  Press 

No

 to use the stored reference weight and proceed to step 6.  

4.  Press 

Yes

 to establish a new reference. Balance will now display [

Put.ref

]. 

5.  Add the desired reference material to the pan or container. Press 

Yes

 to store the 

  reference weight. The display shows 100%.  

6.  Replace the reference material with the sample material. The display shows the 

  percentage of the sample compared to reference weight.  

7.   To clear the stored reference press and hold 

Mode

 until [

Percnt

] is displayed. Press 

Yes

 

when [

Clr.ref

] is displayed. 

Note: Press 

Function

 to view the current reference weight.

 

 
 
 
 
 
 
 

(866) 433-6682 • (281) 359-8538 • [email protected] • www.novatech-usa.com

Summary of Contents for Scout SPX1202

Page 1: ...Scout TM Series Balances SPX Instruction Manual sales novatech usa com www novatech usa com Tel 866 433 6682 Fax 866 433 6684 Tel 281 359 8538 Fax 281 359 0084...

Page 2: ...comply with these warnings could result in personal injury and or property damage Retain all instructions for future reference Verify that the local AC power supply is within the input voltage range p...

Page 3: ...ocation For best performance the Scout SPX balance should be used in a clean stable environment Do not use the balance in environments with excessive drafts with rapid temperature changes near magneti...

Page 4: ...mpartment Note After power on it is recommended to warm up the balance for at least 5 minutes before using it 2 5 Initial Calibration When the Balance is first installed and when it is moved to anothe...

Page 5: ...o Off Turns the balance Off Units Changes the weighing unit Mode Allows changing the application mode Menu Enter the User menu Menu Function Short Press Yes Accepts the current setting on the display...

Page 6: ...nd hold the On Zero Off button until OFF is displayed 3 3 Calibration Operation When the balance is operated for the first time a span calibration is recommended to ensure accurate weighing results Be...

Page 7: ...al pieces until the desired count is reached 8 To clear the stored APW press and hold Mode until Count is displayed Press Yes when Clr pwW is displayed Note Press Function to view the current APW 3 6...

Page 8: ...der icon will light If the sample is within the target weight range the accept symbol will light If the sample is over the target weight range over icon will light Note Press Function to view the low...

Page 9: ...clear the stored total press and hold Mode until total is displayed When Clr tot is displayed press Yes 3 9 Hold Mode There are two modes for the display hold Peak Hold allows the user to capture and...

Page 10: ...Peak Off Reset g kg ct N oz ozt dwt lb lb oz grn hkt sgt twt t tola tical C End End End End Notes Only available in certain regions When the Security Switch is in locked position the Calibration funct...

Page 11: ...hange of 3 divisions per second has been exceeded Stable Range Stable Set the amount the reading can vary while the stability symbol remains on 0 5d 0 5 balance division 1d 1 balance division 2d 2 bal...

Page 12: ...available for use with the Mode button Reset reset Reset the Mode menu to factory defaults NO not reset YES reset Weigh wWEIGH Set the status OFF disabled ON enabled Count COUNt Set the status OFF di...

Page 13: ...ts in the menu must be turned on to be active Note Available units vary by model and local regulations Custom Unit C Use the Custom Unit to display weight in an alternative unit of measure The custom...

Page 14: ...2 rESEt will be displayed Press No Unit g will be displayed 3 Press No to toggle through active units until Unit c is displayed 4 Press Yes On c will be displayed then press Yes 5 Factor is displayed...

Page 15: ...is level and secure and that the transportation lock has been released Power on the balance then use a string or wire to attach items to be weighed Figure 4 3 Weigh below application Connecting the I...

Page 16: ...ght 1d Balance shows error then exits parts counting or goes to Clr APU Lo ref Percent Reference weight or APW is too low for accurate results CAL E Fail to do calibration USB Err Cannot find menu or...

Page 17: ...ration Digital with external weight Tare Range Full Capacity by subtraction Weighing Units g kg ct N oz ozt dwt lb lb oz grn Tael HongKong Tael Singapore Tael Taiwan tical tola C Custom unit Applicati...

Page 18: ...200 g 300 g 2 kg 5 kg 8 kg Linearity Calibration Mass 200 400 g 300 600 g 1 kg 2 kg 3 kg 6 kg 4 kg 8 kg Stabilization Time s 1 Construction ABS plastic housing with 304 stainless steel SST pan Drafts...

Page 19: ...0 0005 4 8500 x 0 0005 13 6685 x 0 0005 18 080 x 0 005 Pound Ounce lb oz 1lb 5 870oz x 0 005oz 4lb 13 600oz x 0 005oz 13lb 10 700oz x 0 005oz 18lb 1 25oz x 0 05oz Grain grn 6480 x 2 9570 x 2 33950 x...

Page 20: ...or electrical and electronic equipment If you have any questions please contact the responsible authority or the distributor from which you purchased this device Should this device be passed on to oth...

Page 21: ...es to be defective at no charge provided that the product is returned freight prepaid to OHAUS This warranty does not apply if the product has been damaged by accident or misuse exposed to radioactive...

Page 22: ...ltweite Gesh ftsstellen Con uffici in tutto il mondo www ohaus com 30268991 P N 30268991 C 2016 OHAUS Corporation all rights reserved todos los derechos reservados tous droits r serv s Alle Rechte vor...

Reviews: