4. MENU SETTINGS

 

The User Menu allows the customizing of balance settings.  
Note: Additional Sub-Menus may be available if Interface Options are installed. See Interface 
User Manual for the additional setting information.

 

 

4.1 Menu Navigation 

User Menu: 

Sub-Menus 

C.a.l 

S.e.t.u.p

 

M.o.d.e 

U.n.i.t

 

E.n.d

 

User Menu 

Items: 

 

Span 
Lin 
 
 
 
 
 
 
 
 
 
 
 
 

 

Reset 

Yes/No 

Filter  

Low/Med/High 

AZT 

0.5/1/3/Off 

StableRange 

0.5/1/2//5 

Backlight 

Off/On/Auto 

Auto Tare 

Off/On/On-acc 

Auto Off 

Off/1/5/10 

 
 

Reset 

Yes/No 

Weighing 

On/Off 

Counting 

On/Off 

Percent 

On/Off 

Check 

On/Off 

Totalize 

On/Off 

Hold 

Disp/Peak/Off 

 
 

Reset 


kg 
ct 

oz 
ozt 
dwt 
lb 
lb:oz 
grn 
hkt* 
sgt* 
twt* 
t*     

 tola/tical 

 

 

End 

End 

End 

End 

 

 

Notes:

   

* Only available in certain regions. 

 

 

When the Security Switch is in locked position, the Calibration function is hidden, Units are 
locked to the current setting. 

 

 

Security switch:

 

 

 

 

4.2 Calibration Menu

 

Enter this menu to perform calibrations.

 

 
 

 

Span

 [

SpaN

]   

Initiates a span calibration procedure (zero and span) 

 

Lin

 [

LIN

]  

Initiates a linearity calibration procedure (zero, mid-point and span). 

 

End Cal

 [

End

Advance to the next menu or return to the top of the current menu.

 

Span:  

Perform 

Linearity: 

Perform 

End Calibration: 

Exit menu 

 

.

 

No/Back 

Long press -

Enter menu

 

Short press -

Accept

 

Short press – 
Go to 

Next/Prev.

 

menu

 

Yes 

Exit 

Short press – 

Exit

 menu

 

(866) 433-6682 • (281) 359-8538 • [email protected] • www.novatech-usa.com

Summary of Contents for Scout SPX1202

Page 1: ...Scout TM Series Balances SPX Instruction Manual sales novatech usa com www novatech usa com Tel 866 433 6682 Fax 866 433 6684 Tel 281 359 8538 Fax 281 359 0084...

Page 2: ...comply with these warnings could result in personal injury and or property damage Retain all instructions for future reference Verify that the local AC power supply is within the input voltage range p...

Page 3: ...ocation For best performance the Scout SPX balance should be used in a clean stable environment Do not use the balance in environments with excessive drafts with rapid temperature changes near magneti...

Page 4: ...mpartment Note After power on it is recommended to warm up the balance for at least 5 minutes before using it 2 5 Initial Calibration When the Balance is first installed and when it is moved to anothe...

Page 5: ...o Off Turns the balance Off Units Changes the weighing unit Mode Allows changing the application mode Menu Enter the User menu Menu Function Short Press Yes Accepts the current setting on the display...

Page 6: ...nd hold the On Zero Off button until OFF is displayed 3 3 Calibration Operation When the balance is operated for the first time a span calibration is recommended to ensure accurate weighing results Be...

Page 7: ...al pieces until the desired count is reached 8 To clear the stored APW press and hold Mode until Count is displayed Press Yes when Clr pwW is displayed Note Press Function to view the current APW 3 6...

Page 8: ...der icon will light If the sample is within the target weight range the accept symbol will light If the sample is over the target weight range over icon will light Note Press Function to view the low...

Page 9: ...clear the stored total press and hold Mode until total is displayed When Clr tot is displayed press Yes 3 9 Hold Mode There are two modes for the display hold Peak Hold allows the user to capture and...

Page 10: ...Peak Off Reset g kg ct N oz ozt dwt lb lb oz grn hkt sgt twt t tola tical C End End End End Notes Only available in certain regions When the Security Switch is in locked position the Calibration funct...

Page 11: ...hange of 3 divisions per second has been exceeded Stable Range Stable Set the amount the reading can vary while the stability symbol remains on 0 5d 0 5 balance division 1d 1 balance division 2d 2 bal...

Page 12: ...available for use with the Mode button Reset reset Reset the Mode menu to factory defaults NO not reset YES reset Weigh wWEIGH Set the status OFF disabled ON enabled Count COUNt Set the status OFF di...

Page 13: ...ts in the menu must be turned on to be active Note Available units vary by model and local regulations Custom Unit C Use the Custom Unit to display weight in an alternative unit of measure The custom...

Page 14: ...2 rESEt will be displayed Press No Unit g will be displayed 3 Press No to toggle through active units until Unit c is displayed 4 Press Yes On c will be displayed then press Yes 5 Factor is displayed...

Page 15: ...is level and secure and that the transportation lock has been released Power on the balance then use a string or wire to attach items to be weighed Figure 4 3 Weigh below application Connecting the I...

Page 16: ...ght 1d Balance shows error then exits parts counting or goes to Clr APU Lo ref Percent Reference weight or APW is too low for accurate results CAL E Fail to do calibration USB Err Cannot find menu or...

Page 17: ...ration Digital with external weight Tare Range Full Capacity by subtraction Weighing Units g kg ct N oz ozt dwt lb lb oz grn Tael HongKong Tael Singapore Tael Taiwan tical tola C Custom unit Applicati...

Page 18: ...200 g 300 g 2 kg 5 kg 8 kg Linearity Calibration Mass 200 400 g 300 600 g 1 kg 2 kg 3 kg 6 kg 4 kg 8 kg Stabilization Time s 1 Construction ABS plastic housing with 304 stainless steel SST pan Drafts...

Page 19: ...0 0005 4 8500 x 0 0005 13 6685 x 0 0005 18 080 x 0 005 Pound Ounce lb oz 1lb 5 870oz x 0 005oz 4lb 13 600oz x 0 005oz 13lb 10 700oz x 0 005oz 18lb 1 25oz x 0 05oz Grain grn 6480 x 2 9570 x 2 33950 x...

Page 20: ...or electrical and electronic equipment If you have any questions please contact the responsible authority or the distributor from which you purchased this device Should this device be passed on to oth...

Page 21: ...es to be defective at no charge provided that the product is returned freight prepaid to OHAUS This warranty does not apply if the product has been damaged by accident or misuse exposed to radioactive...

Page 22: ...ltweite Gesh ftsstellen Con uffici in tutto il mondo www ohaus com 30268991 P N 30268991 C 2016 OHAUS Corporation all rights reserved todos los derechos reservados tous droits r serv s Alle Rechte vor...

Reviews: