Sapphire JS10 and JS500 EN-9
1. When the balance displays
, press
On
.
.
.
.
.
will be displayed
followed by the value of the mass which must be placed on the pan.
Do not disturb the balance while
is displayed.
2. Place the required mass on the pan and momentarily press
On
.
is
displayed, then the value of the second mass to be placed on the pan is
displayed.
3. Place the required mass on the pan and momentarily press
On
.....
4. When the weight on the pan is displayed with the stability indicator, the
balance is calibrated.
When Linearity calibration is completed, the balance automatically exits the menu
and stores any changes you have made.
CARE AND MAINTENANCE
To keep the balance operating properly, keep the housing and platform clean. If
necessary, a cloth dampened with a mild detergent may be used. Unplug the AC
Adapter when not in use. For long term storage, remove the battery.
TROUBLESHOOTING
SYMPTOM
No Display.
BAT
Indicator is on.
Incorrect weight reading.
Calibration procedure
does not work.
Balance won't store Auto-
Shut-off or unit selections
made in menu.
PROBABLE CAUSE
Power Adapter not connected.
Battery is dead.
Battery is weak.
Balance out of calibration.
Balance was not rezeroed before
weighing.
Incorrect calibration masses
being used.
END selection was not used to
exit menu.
REMEDY
Connect AC Adapter.
Replace battery.
Replace battery.
Calibrate the balance.
Press
On
with no weight on the
pan, then weigh item.
Use correct masses.
You must use END to exit menu
and save selections.