Error Codes
Err 0
Invalid checksum data. The balance may need recalibration, particu-
larly linearity calibration. If the error persists after recalibration, the
balance must be serviced.
Err 1
Sample being weighed exceeds the capacity of the balance. If error
occurs when the sample is within the balance capacity, balance may
be incorrectly calibrated. An underload such as the pan off of the
balance could also display Err1. Recalibrate the balance.
Err 4
Incorrect or no calibration mass used when performing calibration
procedure. Error will flash momentarily, then balance will use previous
calibration data. Recalibrate correctly.
Err 9
Internal data error. Return balance for servicing.