Ranger
™
4000 Count Series EN 19
Data Transfer
[
data.tr
]
Output weighing results directly to a PC application.
OFF = disabled
ON
= enabled
Windows XP Setup:
1. Click Start Menu in Windows XP and click Settings -> open
Control Panel
.
2. Double click
Accessibility Options
in Control Panel.
3. Select the
General
tab.
4. Check
Use Serial Keys
, and click the
Settings
button.
5. Select the
Serial Port
, set the Baud rate to
9600
and click
OK
.
6. Close the Control Panel.
7. Run Excel to open one blank sheet. Focus the cursor on one item. At this time, if scale
sends data to the PC through the RS232 port, the data will be put into the cell, and the
cursor will automatically move to the next vertical cell.
Notes:
For Windows 7 setup, contact Ohaus.
If the weighing value is a negative number, set the target cell in TEXT format. Otherwise, Excel
will not distinguish it as a negative number.
Please do not use this function during continuous printing.
Print Library
[
P.lib
]
Print the library data.
No = disabled
Yes = All library records stored in memory are printed
End Print
[
End
]
Advance to the next menu or return to the top of the current menu.
4.7 COM1 Menu
Enter this menu to define communication parameters
Note:
The COM2 menu is only displayed if a second interface is installed.
Reset
[
reset
]
Reset the COM menu to factory defaults.
NO
= no reset
YES = reset
Baud Rate
[
BAUD
]
Set the baud rate.
300
= 300 bps
600
= 600 bps
Reset:
no
, yes
Baud Rate:
300…
9600
…19200
Parity:
7 even, 7 odd, 7 none,
8 none
Stop bit:
1
, 2
Handshake:
none
, On-Off
Alternate command:
Print (A…
P
…Z), Tare (A…
T
…Z), Zero (A…
Z
)
End COM:
Exit menu