Filter
[
FILtEr
]
Set the amount of signal filtering.
LOW
= less stability, faster stabilization time
MED
= normal stability, stabilization time
HI
= greater stability, slower stabilization time
AZT
[
AZt
]
Set the automatic zero tracking functionality.
OFF
= disabled
0.5d
= the display will maintain zero until a change of 0.5 divisions per second has
been exceeded.
1d
= the display will maintain zero until a change of 1 divisions per second has been
exceeded.
3d
= the display will maintain zero until a change of 3 divisions per second has been
exceeded.
Light
[
LIGHt
]
Sets backlight functionality.
OFF
= always off
ON
= always on
AUTO
= turns on when a button is pressed or the displayed weight changes.
Auto off
[
A.OFF
]
Set the automatic shut off functionality.
OFF
= disabled
1
= powers off after 1 minute of no activity
5
= powers off after 5 minute of no activity
10
= powers off after 10 minute of no activity
End Readout
[
End
]
Advance to the next menu or return to the top of the current menu.
4.5 Unit Menu
This sub-menu activates units so they will be accessible with the
Units
button. The units in the
menu must be turned “on” to be active.
Note: Available units vary by model and local regulations.
4.6 Print1 Menu
Enter this menu to set printing parameters. Default settings are
bold
.
Note:
The Print2 menu is only displayed if a second interface is installed.
Reset:
no
, yes
Stable Only:
off
, on
Auto Print:
off
, on stable, interval, continuous, accept
Content:
P/N (->
off
, on)
Result (-> off,
on
)
Gross (->
off
, on)
Net (->
off
, on)
Tare (->
off
, on)
Header (->
off
, on)
Footer (->
off
, on)
Mode (->
off
, on)
Unit (-> off,
on
)
Info (->
off
, on)
APW (->
off
, on)
Acc (->
off
, result, all)
Layout:
Format (->
S