EN 9 Ranger
™ 4000 Series
5. The scale will display [
CL.lImM
] (clear Check limits).
6. Press
No
to use the stored check limits and proceed to step 9.
7. Press
Yes
to establish new check limits. The scale will then display [
Set. Lo
]. Press
Yes
to view the “Low” limit value. Press
Yes
to accept or
No
to edit the “Low” limit value. The
stored value then displays with the first digit highlighted [
0
00000
Pcs]. Repeatedly press
No
until the desired number appears. Press
Yes
to accept and highlight the next digit.
Repeat until all the digits are correct. Press
Yes
to accept the “low” limit value, [
Set. Hi
]
will be displayed.
8. Repeat the same procedure to accept or edit t
he “high” value.
9. If required, place an empty container on the pan and press
Tare
.
10. Place sample material on the pan or in the container. If the sample weight is under the
target quantity range, the yellow LED will light. If the sample is within the target quantity
range, the green LED will light. If the sample is over the target quantity range, the red
LED will light.
Note: Press
Function
to view the low and high check limits.
Positive Check
Positive check is used to determine when the material added to the scale is within the target
range. In this case the UNDER and OVER limits must be positive values. (The OVER limit must
be greater than the UNDER limit.)
Add material to the scale until it is within the ACCEPT (green) range.
Negative Check
Negative check is used to determine when the material removed from the scale is within the
target range. In this case the UNDER and OVER limits are both negative values.
(The UNDER limit must be greater than the OVER limit.)
Place the item to be weighed on the scale and press
TARE
.
Remove a portion of the item until it is within the ACCEPT range.
Zero Check
Zero check is used when comparing subsequent samples to an initial reference sample. In this
case, the UNDER limit must be a negative value and the OVER limit must be a positive value.
Place the reference item on the scale and press
TARE
. Remove the reference sample and
place the item to be compared on the scale to determine if it is within the ACCEPT range.
3.7 Dynamic Mode
This mode allows the user to weigh an unstable load such as a moving animal. The weight is
held on the display until reset. Manual, semi-automatic and automatic start/reset methods are
available.
Manual operation
(DYNAMIC is set to MANUAL in the Mode menu):
1. Press and hold
Mode
until [
dYNAmM
] (Dynamic) is displayed. [
rEAdY
] (Ready) will then
be displayed.
2. Place the load on the pan and press the
Function
key to start measurement.
During the average period, the countdown timer decreases in one second increments.
3. After completed countdown, the average weight will be displayed. The tilde symbol will
blink indicating that the current weight is being held.