Pioneer
™
EN-4
accuracy of the count. The balance automatically recalculates the Average Piece Weight when the number of parts
on the pan is less than three times the original sample size. The display shows [
APwW.OPT
] each time the APW is
optimized.
3.3 Percent Mode
–
Use this mode to measure the weight of a sample as a percentage of a reference weight.
To access the Percent Mode, press
Unit
until the display shows [
PErCEnt
].
Establish a new Reference Weight - With Clear reference [
CLr.rEF
] displayed, Press
No
to use the previously
saved Reference Weight. Press
Yes
to establish a new Reference Weight. Put the reference sample on the pan
and press Yes to accept or No to abort.
Percent – Place the object(s) to be compared to the reference weight on the pan.
3.4
Weigh Below Feature
Remove Weigh Below Cover
Attach wire or string to Hook
Suspend sample
4. SETTINGS
4.1 Menu Navigation
.CAL.
InCal*
SPAN
LINEAR
**
.SETUP.
AZSM
FILTER
A.TARE
GRAD
GBL RST
END.SET
.MODE.
COUNT
APWOPT
PERCENT
END.MOD
.UNIT.
END.UNIT
GRAM
MILLIGRAM
KILOGRAM
OUNCE
NOT ALL
UNITS
SHOWN
.PRINT.
OUT STA
A. PRINT
CONTENT
LINE FO
END.PRT
.RS232.
BAUD
PARITY
HAND.SH
END.R232
.LOC.
CAL
SETUP
MODE
RS232
END.LOCK
.END.
NO
YES
UNIT
POUND
CAL ADJ
* InCal models only
** Not available in InCal models
END.CAL
Enter Menu – When the balance is on, press and hold Unit/Menu until [
mMENU
] appears.
Release the button and the Calibrate [
.CAL.
] menu will display.
Menu Navigation – Select menus, menu items and settings through use of the
Yes
and
No
buttons.
Solid arrows point to the content displayed when
Yes
is pressed, Dashed lines when No is pressed.
Changing Settings – To select the displayed Setting, press
Yes
. To move to the next Setting, press
No
.
Exit Menu – When [
.END
.
] is displayed, press
Yes
to exit the menu function, or press No to return to the Cal menu.
Note:
Press and hold
No
at any time to exit quickly.