EN-9
Pioneer
™
6.2 Specifications
*
C= Internal calibration
6.3 Communication
6.3.1 Commands
The RS232 Interface allows a computer to control the balance as well as receiving data such as the displayed
weight. The balance will return “ES” for invalid commands.
Command Function
IP
Immediate Print of displayed weight.
P
Print displayed weight (uses Stable ON/OFF menu settings).
CP
Continuous Print.
xP
Interval Print x = Print Interval (1-3600 sec)
T
Same as pressing Zero Key.
ON
Turns balance ON.
OFF
Turns balance OFF.
PSN
Show Serial Number.
PV
Version: Print product name, software revision and LFT ON (if LFT is set ON).
PU
Print current mode/unit
x#
Set PC ref wt (x) in grams
P#
Print PC ref wt
x%
Set % ref wt (x) in grams
P%
Print % ref wt
6.3.2 RS232 (DB9) Pin Connections
Pin 2: Balance transmit line (TxD)
Pin 3: Balance receive line (RxD)
Pin 5: Ground signal (GND)
Pin 7: Clear to send (hardware handshake) (CTS)
Pin 8: Request to send (hardware handshake) (RTS)
MODEL
PA64
*
PA114
*
PA214
*
PA213
*
PA413
*
PA512
*
PA2102
*
PA4102
*
PA4101
*
Capacity (g)
65
110
210
210
410
510
2100
4100
4100
Readability (g)
0.0001
0.001
0.01
0.1
Repeatability (sd)
0.0001
0.001
0.01
0.1
Linearity (g)
± 0.0002
± 0.0003
± 0.002
± 0.02
± 0.1
Tare Range
To capacity by subtraction
Stabilization
3 seconds
Span cal mass (g)
50 or
60
50 or
100
100 or
200
100 or
200
200 or
400
200 or
500
1000 or
2000
2000 or
4000
2000 or
4000
Lin. cal masses (g)
25,
50
50,
100
100,
200
100,
200
200,
400
250,
500
1000,
2000
2000,
4000
2000,
4000
Pan size (in / cm)
3.5 / 9 dia.
4.7 / 12 dia.
7.1 / 18 dia.
Net Weight (lb/kg)
10.1 / 4.6 (
*
11.2 / 5.2)
10 / 4.5 (
*
11.1 / 5.1)
CTS
RTS
5
9
1
6
RxD
TxD
GND