TABLE 9-5. SPECIFICATIONS (continued)
InCal Approval Model
3;0
3;0
3;0
Capacity (g)
3200
5200
4200
Readability d (g)
0.01
0.01
0.1
Verification Interval e (g)
0.1
0.1
0.1
Class
II
II
II
Repeatability (STDEV) (g)
0.01
0.01
0.1
Linearity (g)
±0.02
±0.02
±0.2
Stabilization Time Typical (s)
1
1
1
Sensitivity temperature drift
(PPM/K)
±3
±3
±10
Typical Minimum Weight USP
(USP K=2,U=0.10%)
20 g
20 g
200 g
Optimized Min-Weight (g) (USP,
u=0.10%, k=2) SRP
0.41d*
8.2 g
8.2 g
82 g
Units
g, kg , ct
Applications
Basic Weighing, Parts Counting, Percent Weighing, Dynamic Weighing,
Density Determination
Platform Size (diameter, mm)
180
180
180
Span Calibration Points (g)
1500, 3000
2500, 5000
2000, 4000
Linearity Calibration Points (g)
0, 1500, 3000
0, 2500, 5000
0, 2000, 4000
Tare Range
To capacity by subtraction
Power Supply
Power input: 100-240V ~ 200mA 50-60Hz 12-18VA
Power output
12 VDC 0.5A
Assembled Dimensions
(W x D x H) (mm)
209 x 321 x 98
Communication
RS232, USB
Operating Temperature Range
Operating conditions for ordinary lab application: +10°C to 30°C (operability
guaranteed b5°C and 40°C).
Storage Temperature Range
Humidity: maximum relative humidity 80% for temperatures up to 30°C,
decreasing linearly to 50% relative humidity at 40°C
Storage Conditions
-10°C to 60°C, humidity 10% to 90%, without condensation
Net Weight
7.7 lb / 3.5 kg
Shipping Weight
11 lb / 5 kg
Shipping Dimensions
(W x D x H) (mm)
550 x 385 x 291
1RWH