3;6HULHV%DODQFH
(1
0HDVXULQJWKH'HQVLW\RID6LQNLQJ6ROLG8VLQJ:DWHU
Press and hold the
)XQFWLRQ 0RGH
button until
Density
appears. Press
<HV
to initiate the
Density Determination
.
,WHP6HWWLQJV
x
Sample type: Solid
x
Auxiliary
Liquid:
Water
x
Porous
Material:
Off
x
Water Temperature: Measure the actual water
temperature using a precision thermometer.
The water temperature is 20.0
đ
by default.
Press
1R
or
%DFN
to increase or decrease the value of
temperature. The balance calculates water density based on
the water temperature value entered.
1.
Weigh the sample in air using the balance and the density
determination kit.
When the * (symbol of stability) appears, press the
)XQFWLRQ 0RGH
button to confirm the weight of the
sample in the air.
2.
Weigh the sample submerged in the liquid using the
balance and the density determination kit.
1RWH
Lower the sample down into the liquid until it is fully
submerged.