OHAUS Navigator NV1101 Instruction Manual Download Page 11

Navigator

®

 

 

EN-9

 

 

6.1 Specifications 

Typical specifications: 

Stabilization Time 

1 seconds 

Tare range 

To capacity by subtraction 

Application Modes

1

 

weigh, percent, parts count, check weigh, accumulate, hold 

Weighing Units 

Non-Approved: ct, g, kg, grn, lb, oz, lb:oz, ozt, N, dwt, tael (3), tola, tical 
M models: g, kg, ct        N models: g, kg, ct, lb, oz, ozt, dwt, grn 

AC Power 

AC Adapter (supplied)- 12 VDC or 12VAC, 420 mA  

Battery Power 

NV: 4 AA (LR6) batteries (not supplied) 
NVL, NVT: 4 C (LR14) batteries (not supplied) or rechargeable battery option 

Calibration 

Digital with external weight 

LCD Display  

6-digit 7-segment LCD with white LED backlight 

LED Display 

6-digit 7-segment LED 

Display Size 

20 mm / 0.78” digits 

Keypad 

4-button overlay plus two programmable IR sensors 

Ingress Protection 

IP43 (NVT and NVL models) 

Pan Size (W x D) 

NV: 190 mm x 138 mm      NV212: Ø120mm 
NVL: 194 mm x 203 mm    NVT: 230 mm x 174 mm 

Net Weight

2

 

NV: 0.9 kg / 2.2 lb      NVL: 1.5 kg / 3.3 lb     NVT: 1.5 kg / 3.3 lb 

Shipping Weight

2

 

NV: 1.4 kg / 3.1 lb      NVL: 2.3 kg / 5.1 lb     NVT: 2.3 kg / 5.1 lb 

Note 1: Some application modes are restricted in Approved models. 
Note 2: /1 and /3 models add 0.5kg / 1lb 
 
Non-Approved Models: 

Model

1

 Capacity 

Readability 

(d)  Repeatability Linearity  Span Cal. Wt.  Lin. Cal. Wts. 

NV212 

210g 0.01g 

1d  ±2d  200g 100g, 

200g 

NV511 

510g 0.1g 

1d  ±2d 300g 300g, 

500g 

NV1101 

1100g 0.1g 

1d  ±2d  500g 500g, 

1000g 

NV2101 

2100g 0.1g 

1d  ±2d  1kg  1kg, 

2kg 

NV4101 

4100g 0.2g 

1d  ±2d  2kg  2kg, 

4kg 

NV5101 

5100g 0.5g 

1d  ±2d  3kg  3kg, 

5kg 

NVL511 

510g 0.1g 

1d  ±2d 300g 300g, 

500g 

NVL1101 

1100g 0.1g 

1d  ±2d  500g 500g, 

1000g 

NVL2101 

2100g 0.1g 

1d  ±2d  1kg  1kg, 

2kg 

NVL5101 

5100g 0.5g 

1d  ±2d  3kg  3kg, 

5kg 

NVL10000 

10000g 1g 

1d  ±2d  5kg  5kg, 

10kg 

NVL20000 

20000g 1g 

1d  ±2d  10kg 10kg, 

20kg 

NVT1601 1600g  0.1g 

1d 

±2d 

1kg 

1kg,1.5kg 

NVT3201 3200g  0.2g 

1d 

±2d 

2kg 

2kg, 

3kg 

NVT6401 6400g  0.5g 

1d 

±2d 

5kg 

3kg, 

6kg 

NVT10001 10000g 

0.5g 

1d 

±2d 

5kg 

5kg, 

10kg 

NVT16000 16000g 

1g 

1d 

±2d 

10kg 

10kg, 

15kg 

Summary of Contents for Navigator NV1101

Page 1: ...Navigator Balances Instruction Manual 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com ...

Page 2: ......

Page 3: ...t Press in Menu Selects accepts displayed setting Print Units No Short Press See Interface Manual for operation description Long Press Toggles through active units Short Press in Menu Toggles through available settings Function Mode Back Short Press Selects function setting Long Press Selects active Mode Short Press in Menu returns to previous settings Tare Menu Exit Short Press Enter clear a Tare...

Page 4: ...nsportation Lock The Transportation Lock is located under the balance Rotate the pointer to the unlocked position Note No shipping lock on Approved models 2 3 Location Use the balance on a firm steady surface Avoid locations with excessive air current vibrations heat sources or rapid temperature changes Adjust the leveling feet so the bubble is centered in the circle ...

Page 5: ...ion manual for step by step instructions as well as disposal instructions Caution Risk of explosion can occur if the rechargeable battery is replaced with the wrong type or if it is not properly connected 2 5 Initial Calibration When the balance is first installed it should be calibrated to ensure accurate results Press and hold Menu until mMeNU Menu is displayed When the button is released the di...

Page 6: ...Yes to calculate the APW The display shows the piece count Note Press Function to view the current APW 6 Add additional pieces until the desired count is reached 7 To clear the stored APW press and hold Mode until Count is displayed Press Yes when Clr APU is displayed 3 3 Percent Mode This mode measures the weight of a sample as a percentage of a reference weight 1 Place an empty container on the ...

Page 7: ... 5 for all of the items to be added Note While the display indicates zero press Function to view the current total number of samples and the total weight 7 To clear the stored total press and hold Mode until ACCUmM is displayed Press Yes when Clr Acc is displayed 3 6 Checkweigh Mode This mode sets low and high weight limits for portion control processes 1 Press and hold Mode until CHeCk Check is d...

Page 8: ... menu 4 2 Cal Sub Menu Span SpaN yes no Initiates a span calibration procedure zero and span A span calibration is important when initially setting up the balance Lin Lin yes no Initiates a linearity calibration procedure zero mid point and span 4 3 Setup Sub Menu The backlit LCD and red LED displays will have different menu items or settings based on the functionality Auto Off A Off on off When A...

Page 9: ...it menus The switch in type approved models may set some balance settings as required by the approval agency The switch may be secured using paper seals wire seals or plastic ties Locked Unlocked 5 MAINTENANCE 5 1 Troubleshooting The following table lists common problems and possible causes and remedies If the problem persists contact OHAUS or your authorized dealer Symptom Possible Cause Remedy C...

Page 10: ...659 between 8 00 AM and 5 00 PM EST An OHAUS product service specialist will be available to provide assistance Outside the USA please visit our web site www ohaus com to locate the OHAUS office nearest you 5 3 Accessories Rechargeable Battery Kit 83032106 NVL NVT RS232 Interface Kit 83032107 NV NVL NVT USB Interface Kit 83032108 NV NVL NVT Ethernet Interface Kit 83032109 NV NVL NVT In Use Cover K...

Page 11: ...L 194 mm x 203 mm NVT 230 mm x 174 mm Net Weight2 NV 0 9 kg 2 2 lb NVL 1 5 kg 3 3 lb NVT 1 5 kg 3 3 lb Shipping Weight2 NV 1 4 kg 3 1 lb NVL 2 3 kg 5 1 lb NVT 2 3 kg 5 1 lb Note 1 Some application modes are restricted in Approved models Note 2 1 and 3 models add 0 5kg 1lb Non Approved Models Model1 Capacity Readability d Repeatability Linearity Span Cal Wt Lin Cal Wts NV212 210g 0 01g 1d 2d 200g 1...

Page 12: ... 0 5g III 1kg 1kg 1 5kg NVT3200 3200g 1g III 2kg 2kg 3kg NVT6400 6400g 2g III 5kg 3kg 6kg NVT16000 16000g 5g III 10kg 10kg 15kg Note 1 M EC Type approved N NTEP and Measurement Canada approved Note 2 1 Factory installed rechargeable battery option 2 LED display 3 LED display with factory installed rechargeable battery option Example NVT3200M 3 6 2 Drawings A B C NV 204 mm 8 in 212 mm 8 4 in 58 mm ...

Page 13: ...dopted European standard EN 45501 1992 paragraph 8 2 2 If national regulations limit the validity period of the verification the user of the weighing instrument must strictly observe the re verification period and inform the respective weights and measures authorities Disposal In conformance with the European Directive 2002 96 EC on Waste Electrical and Electronic Equipment WEEE this device may no...

Page 14: ...was re registered to the ISO 9001 2000 standard Limited Warranty OHAUS products are warranted against defects in materials and workmanship from the date of delivery through the duration of the warranty period During the warranty period OHAUS will repair or at its option replace any component s that proves to be defective at no charge provided that the product is returned freight prepaid to OHAUS T...

Reviews: