EN-8

 

Navigator

®

 

 

Err 8.1 

 

¯

Load

¯

 

Power on zero range exceeded 

Clear pan, check Shipping Lock setting 

Err 8.2  _Load_

 

Power on zero under range 

Install pan, check Shipping Lock setting 

Err 8.3 

 

¯

Load

¯

 

Overload (>cap+9e) 

Load exceeds balance maximum capacity 

Err 8.4  _Load_

 

Under load 

Reading below min. range - Re-install pan. 

Err 8.6  999999

 

Displayed value >999999  

Result exceeds display capability. 

Err 9   Data

 

Internal data error. 

Contact an authorized service agent 

Err 13 mMEmM

 

Fail to write EEPROM. 

Contact  an authorized service agent 

Err 53  CSumM

 

Invalid checksum data 

Contact  an authorized service agent 

 

5.2 Service Information 

If the troubleshooting section does not resolve or describe your problem, contact your authorized 
OHAUS service agent. For service assistance or technical support in the United States call toll-free 1-
800-526-0659 between 8:00 AM and 5:00 PM EST. An OHAUS product service specialist will be 
available to provide assistance. Outside the USA, please visit our web site, www.ohaus.com to locate 
the OHAUS office nearest you.  
 

5.3 Accessories 

Rechargeable Battery Kit 

83032106 (NVL, NVT) 

RS232 Interface Kit 

83032107 (NV, NVL, NVT) 

USB Interface Kit 

83032108 (NV, NVL, NVT) 

Ethernet Interface Kit 

83032109 (NV, NVL, NVT) 

In-Use Cover Kit 

83032223 (NV) 

83032222 (NVL)  83032221 (NVT) 

Carrying Case Kit 

83032226 (NV) 

83032225 (NVL)  83032224 (NVT) 

Printers and Cables 

Contact OHAUS 

 
 

6. TECHNICAL DATA  

The technical data is valid under the following ambient conditions: 
Ambient temperature: 10°C to 40°C, Approved models (M) 0°C to 40°C 
Relative humidity: 20% to 85% relative humidity, non-condensing 
Height above sea level: Up to 4000 m 
Operability: assured at ambient temperatures between 0°C and 40°C 
Power: AC Adapter – 12VDC 420mA output, 4 batteries or rechargeable battery 
Protection: dust and water 
Pollution degree: 2  
Installation category: Class III 
EMC: See Declaration of Conformity 

Summary of Contents for Navigator NV1101

Page 1: ...Navigator Balances Instruction Manual 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com ...

Page 2: ......

Page 3: ...t Press in Menu Selects accepts displayed setting Print Units No Short Press See Interface Manual for operation description Long Press Toggles through active units Short Press in Menu Toggles through available settings Function Mode Back Short Press Selects function setting Long Press Selects active Mode Short Press in Menu returns to previous settings Tare Menu Exit Short Press Enter clear a Tare...

Page 4: ...nsportation Lock The Transportation Lock is located under the balance Rotate the pointer to the unlocked position Note No shipping lock on Approved models 2 3 Location Use the balance on a firm steady surface Avoid locations with excessive air current vibrations heat sources or rapid temperature changes Adjust the leveling feet so the bubble is centered in the circle ...

Page 5: ...ion manual for step by step instructions as well as disposal instructions Caution Risk of explosion can occur if the rechargeable battery is replaced with the wrong type or if it is not properly connected 2 5 Initial Calibration When the balance is first installed it should be calibrated to ensure accurate results Press and hold Menu until mMeNU Menu is displayed When the button is released the di...

Page 6: ...Yes to calculate the APW The display shows the piece count Note Press Function to view the current APW 6 Add additional pieces until the desired count is reached 7 To clear the stored APW press and hold Mode until Count is displayed Press Yes when Clr APU is displayed 3 3 Percent Mode This mode measures the weight of a sample as a percentage of a reference weight 1 Place an empty container on the ...

Page 7: ... 5 for all of the items to be added Note While the display indicates zero press Function to view the current total number of samples and the total weight 7 To clear the stored total press and hold Mode until ACCUmM is displayed Press Yes when Clr Acc is displayed 3 6 Checkweigh Mode This mode sets low and high weight limits for portion control processes 1 Press and hold Mode until CHeCk Check is d...

Page 8: ... menu 4 2 Cal Sub Menu Span SpaN yes no Initiates a span calibration procedure zero and span A span calibration is important when initially setting up the balance Lin Lin yes no Initiates a linearity calibration procedure zero mid point and span 4 3 Setup Sub Menu The backlit LCD and red LED displays will have different menu items or settings based on the functionality Auto Off A Off on off When A...

Page 9: ...it menus The switch in type approved models may set some balance settings as required by the approval agency The switch may be secured using paper seals wire seals or plastic ties Locked Unlocked 5 MAINTENANCE 5 1 Troubleshooting The following table lists common problems and possible causes and remedies If the problem persists contact OHAUS or your authorized dealer Symptom Possible Cause Remedy C...

Page 10: ...659 between 8 00 AM and 5 00 PM EST An OHAUS product service specialist will be available to provide assistance Outside the USA please visit our web site www ohaus com to locate the OHAUS office nearest you 5 3 Accessories Rechargeable Battery Kit 83032106 NVL NVT RS232 Interface Kit 83032107 NV NVL NVT USB Interface Kit 83032108 NV NVL NVT Ethernet Interface Kit 83032109 NV NVL NVT In Use Cover K...

Page 11: ...L 194 mm x 203 mm NVT 230 mm x 174 mm Net Weight2 NV 0 9 kg 2 2 lb NVL 1 5 kg 3 3 lb NVT 1 5 kg 3 3 lb Shipping Weight2 NV 1 4 kg 3 1 lb NVL 2 3 kg 5 1 lb NVT 2 3 kg 5 1 lb Note 1 Some application modes are restricted in Approved models Note 2 1 and 3 models add 0 5kg 1lb Non Approved Models Model1 Capacity Readability d Repeatability Linearity Span Cal Wt Lin Cal Wts NV212 210g 0 01g 1d 2d 200g 1...

Page 12: ... 0 5g III 1kg 1kg 1 5kg NVT3200 3200g 1g III 2kg 2kg 3kg NVT6400 6400g 2g III 5kg 3kg 6kg NVT16000 16000g 5g III 10kg 10kg 15kg Note 1 M EC Type approved N NTEP and Measurement Canada approved Note 2 1 Factory installed rechargeable battery option 2 LED display 3 LED display with factory installed rechargeable battery option Example NVT3200M 3 6 2 Drawings A B C NV 204 mm 8 in 212 mm 8 4 in 58 mm ...

Page 13: ...dopted European standard EN 45501 1992 paragraph 8 2 2 If national regulations limit the validity period of the verification the user of the weighing instrument must strictly observe the re verification period and inform the respective weights and measures authorities Disposal In conformance with the European Directive 2002 96 EC on Waste Electrical and Electronic Equipment WEEE this device may no...

Page 14: ...was re registered to the ISO 9001 2000 standard Limited Warranty OHAUS products are warranted against defects in materials and workmanship from the date of delivery through the duration of the warranty period During the warranty period OHAUS will repair or at its option replace any component s that proves to be defective at no charge provided that the product is returned freight prepaid to OHAUS T...

Reviews: