![OHAUS Frontier FC5515 Instruction Manual Download Page 39](http://html1.mh-extra.com/html/ohaus/frontier-fc5515/frontier-fc5515_instruction-manual_3560327039.webp)
Frontier
TM
Centrifuge FC5515/FC5515R EN-31
12.2 Table 2: Permissible net weight
Rotor-Number
max speed
permissible
weight
30130870; 24x1.5ml/2.0ml
15200 min-1
82 g
30130871; 24x1.5ml/2.0ml
15200 min-1
82 g
30130872; 30x1.5ml/2.0ml
13500 min-1
102 g
30130879; 44x1.5ml/2.0ml
13500 min-1
150 g
30130874; 4x8place PCR stripes
15000 min-1
14 g
30130873; 12x5ml
14500 min-1
165 g
30130881; 24 Hematocrit
12000 min-1
4.8 g
12.3 Table 3: Lowest temperatures at max. speed
Rotor-Number
max speed
n-max
30130870; 24x1.5ml/2.0ml
15200 min-1
+5 °C
30130871; 24x1.5ml/2.0ml
15200 min-1
+5 °C
30130872; 30x1.5ml/2.0ml
13500 min-1
+6 °C
30130879; 44x1.5ml/2.0ml
13500 min-1
+6 °C
30130874; 4x8place PCR stripes
15000 min-1
+5 °C
30130873; 12x5ml
14500 min-1
+4 °C
30130881; 24 Hematocrit
12000 min-1
Air cool
All temperature indications refer to a room temperature of 23°C. By exceeding this value or direct solar radiation to
the centrifuge, these values can´t be kept up.
12.4 Table 4: Max. speed and RCF-values for permissible rotors
Rotor-Number
max speed
RCF value
30130870; 24x1.5ml/2.0ml
15200 min-1
21953xg
30130871; 24x1.5ml/2.0ml
15200 min-1
21953xg
30130872; 30x1.5ml/2.0ml
13500 min-1
19150xg
30130879; 44x1.5ml/2.0ml
13500 min-1
17113 x g /17317 x g
30130874; 4x8place PCR stripes
15000 min-1
15342xg
30130873; 12x5ml
14500 min-1
19978xg
30130881; 24 Hematocrit
12000 min-1
14970xg
12.5 Table 5: Acceleration and deceleration times
FC5515R
Acceleration values
Deceleration values
Rotor-Number
level 0
level 9
level 0
level 9
30130870; 24x1.5ml/2.0ml
153
20
287
17
30130871; 24x1.5ml/2.0ml
153
20
287
17
30130872; 30x1.5ml/2.0ml
235
29
258
27
30130879; 44x1.5ml/2.0ml
136
18
133
17
30130874; 4x8place PCR stripes
150
17
95
20
30130873; 12x5ml
148
24
320
17
in seconds
Acceleration time
Deceleration time
from 0 min
-1
-> U
max
from U
max
-> 0 min
-1