![OHAUS Frontier FC5515 Instruction Manual Download Page 18](http://html1.mh-extra.com/html/ohaus/frontier-fc5515/frontier-fc5515_instruction-manual_3560327018.webp)
EN-10 Frontier
TM
Centrifuge FC5515/FC5515R
Figure. 18
3.3.4
Pre-selection of temperature (Only FC5515R)
This function is activated by the key
"Temp/Setup"
(13). After pressing this key in the display
"Time/Prog"
the
indication
"°C"
(A-4) flashes. By the adjusting knob (1) the desired test temperature can be pre-selected in steps of
1°C in a range from -20°C up to +40°C.
The value is indicated permanently in the display (figure 19) - before, during and after the run.
Please notice the respective lowest temperatures of the rotors at maximum speed!
Figure. 19
3.3.5
Pre-cooling (Only FC5515R)
If the samples are temperature-sensitive it is useful to pre-cool the centrifuge, the rotor and eventually the buckets
to the required working temperature. Therefore, insert the desired rotor and pre-set the respective temperature. By
simultaneous pressing the keys
"Temp/Setup"
(13) (refer to figure 20) and
"Time"
(6) you start the run. While
running, the unit chooses automatically a rotational speed that is equivalent to 20 % of the permitted rotational
speed of the respective rotor. After the pre-set temperature is reached you can leave the pre-cooling run with the
"Stop"
key (10).
Depending on the inserted rotor the pre-cooling goes between approx. 10 and 20 min.
M8
M7
A-2
1
5
13
1
A-4