EXPLORER BALANCES
EN-47
The Filling Application Setups
(defaults in Bold)
Item Available
Settings
Comments
Enabled
On
, Off
Turns Application on/off
Secondary Unit
On,
Off
To show the Second Display line
Statistics
Off,
Manual, Auto
Lock Settings
On,
Off
Press to lock the current setting(s) so they
cannot be changed.
Reference fields
Target Weight
On
, Off
The Target weight is displayed
Difference (Weight)
On
, Off
Displays the Difference to the Target weight
Difference (Percent)
On
, Off
Displays the Difference to the Target weight
Gross On,
Off
Gross weight is displayed
Net On,
Off
Net weight is displayed
Tare On,
Off
Tare weight is displayed
Print Options
See
Printing
Section.
Target Weight
On
, Off
Difference (Weight)
On
, Off
Difference (Percent)
On
, Off
4.6.3 Statistics for Filling
Use the Statistics function to compare a number of samples and examine the relative deviation of the samples
along with other statistical data. A minimum of three samples is required.
Statistics can be determined manually (
Manual
mode, with key press after each step) or automatically
(
Automatic
Mode: weights are automatically recorded when stable). To use the statistics function, enable it first
in the Application Setup menu.
Application Setup
First, Enable Statistics
Touch the Setup
Icon wrench (see
below) to enter
Setup menu.
Touch
Statistics
.
Choose either
Manual
or
Automatic
mode, then
press
Done
.