Defender 3000 Indicators
EN-35
4.5
Unit Menu
Enter this menu
U.N.I.t
to activate the desired units.
Reset
Gram (g)
Kilogram (kg)
Pound (lb)
Ounce (oz)
Pound:Ounce (lb:oz)
Metric Tonne (t)
End
Note:
Available units vary by model. In addition, due to national laws, the indicator may not include some of
the units listed.
If the security switch is turned on, the Units menu will be locked at their current setting.
When the scale capacity is set to be more than 1000 kg,
Metric
Tonne (t)
unit will appear in the Unit
menu.
4.6
Communication
Enter this menu to define external communication methods and to set printing parameters. Data may be output
to either a printer or PC. Factory default settings are shown in bold.
4.6.1 RS232 Menu
Enter this menu to define communication parameters.
Menu
Sub-Menu
Sub-Menu
(in segment)
Options
Options (in segment)
RS232
(
r.S.2.3.2
)
Baud Rate
BAUd
300, 600, 1200, 2400, 4800,
9600
, 19200, 38400, 57600
/
Parity
PArItY
7 Even, 7 Odd, 7 None,
8
None
7 EuEN
,
7 Odd
,
7
NONe
,
8 NONE
Stop Bit
StOP
1 bit
, 2 bit
1 bIt
,
2 bIt
Handshake
H.SHAKE
None
, Xon/Xoff
NONe
,
ON-OFF
Alt Print CMD
ALt.P
'A' ~ 'Z',
P
/
Alt Tare CMD
ALt.t
'A' ~ 'Z',
T
/
Alt Zero CMD
ALt.Z
'A' ~ 'Z',
Z
/
Reset
reset
no
, yes
NO
,
yes
End
End
/
/
Note
: / refers to not applicable in the table above.
Reset
[
reset
]
Reset the RS232 menu to factory defaults.
NO
= do not reset.
YES
= reset
Baud Rate
[
BAUd
]
Set the baud rate (bits per second).
300
= 300 bps
600
= 600 bps
1200
= 1200 bps
2400
= 2400 bps
4800
= 4800 bps
9600
= 9600 bps
19200
= 19200 bps
38400
= 38400 bps
57600
= 57600 bps