Defender Series Base
EN-2
MAINTENANCE
Cleaning
The base components should be kept clean and free of excessive material build up.
•
Damp cloth with water and a mild detergent may be used to wipe clean the external surfaces—do not use acids,
alkalis, strong solvents or abrasive materials and chemicals
Troubleshooting
Aside from installing components and leveling adjustments, the Defender Series Base does not require any other adjustments as
shipped from the factory.
Operational difficulties that may be encountered can often be traced to simple causes such as:
•
Loose or incorrect wiring connections
•
Obstructions to the base frame
•
Unstable environments
•
Incorrect calibration or setup of the indicator
If the troubleshooting section does not resolve or describe your problem, contact your authorized Ohaus service agent. For
service assistance or technical support in the United States call toll-free 1-800-526-0659 between 8:00 AM and 5:00 PM EST.
An Ohaus product service specialist will be available to provide assistance. Outside the USA, please visit our web site,
www.ohaus.com to locate the Ohaus office nearest you.
TECHNICAL DATA
Technical Specifications
The technical data is valid under the following ambient conditions:
Ambient temperature: -10º C to +40º C
Relative humidity: up to 80% at 31º C, decreasing linearity to 50% at 40º C
Height above sea level: Up to 2,000m
Operability is assured at ambient temperature -10º C to + 40º C
TABLE 2. SPECIFICATIONS
MODEL D30ER
D60ER
D150EL
D150EX
D300EX
Capacity
30 kg
60 kg
150 kg
150 kg
300 kg
Maximum Resolution
3000d
3000d
3000d
3000d
3000d
Pan Dimensions
300 x 350 x 50 mm
400 x 500 x 50 mm
500 x 600 x 100 mm
Base Construction
430 steel platform with painted carbon steel frame and rubber leveling feet
Repeatability (std. deviation)
1d
Linearity ±1d
Load Cell Cable
Length 3m, 4-wire
Length 3m, 6-wire
Load Cell Type
350 Ohm, aluminum, single point
Load Cell Excitation
5-15V DC/AC
Load Cell Rated Output
2mV/V
Load Cell Protection
IP67
Safe Overload Capacity
150% of capacity