2.7 Initial Scale Setup (Cont.)
2.7.7 Lockout Switch Menu
Lockout Switch menu (LOCSW) is a software controlled option which can lock the settings in the Calibration, Setup,
Readout, and Print menus to prevent tampering. When used in conjunction with the Lock Switch (jumper) on the
printed circuit board, the Calibration, Setup, Readout and Print menus can be read only and not changed by an
operator.
Procedure
To select any of the items in the Lockswitch menu, pro-
ceed as follows:
NOTE
: If you have entered from the
preceeding menu, disregard the first step.
• With the Indicator ON, press and hold the
Zero/Menu
button until MENU is displayed. When you release
the
Zero/Menu
button, CAL is displayed, then press
Print/Units
button, until LOCSW is displayed.
• Press
Zero/Menu
button, LSTOFF is displayed. This
permits locking the Setup menu. OFF is unlocked, ON
is read only (locked). This menu is hidden if CAL
jumper is off.
• Press
Print/Units
button, and select ON or OFF.
• Press
Zero/Menu
button, LrdOFF displayed. This
permits locking the Readout menu. OFF is unlocked,
ON is read only (locked).
• Press
Print/Units
button, and select ON or OFF.
• Press
Zero/Menu
button, LPtOFF is displayed. This
permits locking the Print menu. OFF is unlocked, ON
is read only (locked).
• Press
Print/Units
button, and select ON or OFF.
• Press
Zero/Menu
button, LCLOFF is displayed. This
permits locking the Calibration menu. OFF is un
locked, ON is read only (locked). This menu is hid-
den if the CAL jumper is off.
• Press
Print/Units
button, and select ON or OFF.
• Press
Zero/Menu
button to end this block, END is
displayed.
• Press
Zero/Menu
button, Quit is displayed.
• Press
Print/Units
button to go to CAL or press
Zero/
Menu
button, Indicator returns to a weighing mode.
NOTE
: At this point, the Indicator must be calibrated and
the jumper removed from the CAL connector in order to
lock out the menus. The top cover of the Indicator should
be free to gain access to the CAL jumper.
Stable Center
of Zero
Gross
Brutto
CD-11 Indicator
18
CD-11 Indicators are equipped with a standard IBM
TM
compatible, bi-directional RS232 interface for communi-
cation with printers and computers. When the Indicator is
connected directly to a printer or Programmable Logic
Controller (PLC), displayed data can be recorded at any
time by simply pressing the
PRINT/UNITS
button.
Connecting the Indicator to a computer or PLC enables
you to operate several functions of the Indicator from the
computer, as well as receive data such as displayed
weight, weighing mode, stability status, etc.
2.7 Initial Scale Setup (Cont.)
2.7.8 Connecting the RS232 Interface
Hardware
A 9-pin female “D” connector is located on the left side of
the indicator is provided for interfacing to other devices.
Pin connections are shown in the adjacent illustration.
RS-232 Pin Connections.
RS-232 Connector Pin Layout.
1
N/C
2
RXD
3
TXD
4
N/C
5
GND
6
N/C
7
N/C
8
N/C
9
N/C