CATAPULT 5000 SERIES SCALES EN-27
6.
MAINTENANCE
CAUTION: DISCONNECT THE UNIT FROM THE POWER SUPPLY BEFORE CLEANING.
6.1
Cleaning
• The housing may be cleaned with a cloth dampened with a mild detergent if necessary.
• Do not use solvents, chemicals, alcohol, ammonia or abrasives to clean the housing or control panel.
6.2
Troubleshooting
TABLE 7-1. TROUBLESHOOTING.
Error Code
Description
Cause
Error 8.1
Power On Error
Weight reading exceeds Power On Zero limit.
Error 8.2
Power On Error
Weight reading below Power On Zero limit.
Error 8.3
Over Range Error
Weight reading exceeds Overload limit.
Error 8.4
Under Range Error
Weight reading below Underload limit.
Error 8.5
Tare out of range Error
Tared at one unit but after switching to another unit the tare
value exceed the max.
Error 8.6
Display Overflow
Weight exceeds 6 digits. Happened in the cases of
Accumulation or counting PCS display
-------
Busy message
Displayed during tare setting, zero setting, printing
--NO--
Action not allowed message
Function not executed.
Battery icon flashing
Low Battery error
Battery is empty
CAL E
Calibration Error
Calibration value outside allowable limits
Lo.rEF
Low reference weight warning
message
Average Piece Weight too small. (Warning)
rEF.Err
Unacceptable reference
weight message
Reference Weight too small. The weight on the pan is too small
to define a valid reference weight.
6.3
Service Information
If the troubleshooting section does not resolve your problem, contact an authorized Ohaus Service Agent. For
Service assistance in the United States, call toll-free 1-800-526-0659 between 8:00 AM and 5:00 PM Eastern
Standard Time. An Ohaus Product Service Specialist will be available to assist you. Outside the USA, please
visit our website www.ohaus.com to locate the Ohaus office nearest you.
7.
TECHNICAL DATA
7.1
Specifications
The technical data is valid under the following ambient conditions:
Indoor use only
Ambient temperature:
-10°C to 40°C / 14°F to 104°F
Relative humidity:
Maximum relative humidity 80% for temperatures up to 31°C decreasing
linearly to 50% relative humidity at 40°C.
Altitude:
up to 2000m
Mains supply voltage fluctuations: up to ±10% of the nominal voltage
Installation category:
II
Pollution degree:
2
Protection class:
III