CATAPULT 5000 SERIES SCALES                                                                                                          EN-11 

 

3.3.2  Span Calibration

 

Span Calibration uses one point. The span point is established with a calibration mass 
placed on the scale. 
 
When SPAN is displayed, press the Yes button to access the Span Calibration menu 
item. 
 
 
The display flashes the span calibration point. Place the specified weight on the scale 
and press the Yes button. 
 
 
To choose a different span point, repeatedly press the No button to increment the 
selections or press the Back button to decrement the selections. Refer to Table 3-3 for 
available span points. When the desired value is displayed, place the specified weight on 
the scale and press the Yes button. 
 
 
 
The display shows --C-- while the span point is established. 
 
If span calibration was successful, the scale exits to the next Calibration menu and 
displays [

LINEAr

 
Note: Span Calibration should be performed after Zero Calibration. 

 

3.3.3  Linearity Calibration 

 

Linearity calibration uses 3 calibration points. The first calibration point is established 
with no weight on the scale. The second calibration point is established at approximately 
half capacity. The third calibration point is established at capacity. The Linearity 
calibration points are fixed and cannot be altered by the user during the calibration 
procedure. Refer to Table 3-3 for the linearity points. 
 
When 

LINEAr

 is displayed, press the Yes button to access the Linearity Calibration 

menu item. 
The display flashes 0. With no weight on the scale, press the Yes button to establish the 
zero point. 
 
The display shows --C-- while the zero point is established.  
 
The display flashes the mid calibration point.  
 
Place the specified weight on the scale and press the Yes button. 
 
 
The display shows --C-- while the mid point is established. 
 
The display flashes the full calibration point.  
 
Place the specified weight on the scale and press the Yes button. 
 
The display shows --C-- while the full point is established. 
 
If linearity calibration was successful, the scale exits to the next Calibration  
menu and displays [

GEO

]. 

 
 
 
 

Summary of Contents for C51XE100L

Page 1: ...Catapult 5000 Series Scales Instruction Manual...

Page 2: ......

Page 3: ...nal 6 2 2 4 RS232 interface Cable to Indicator Optional 7 2 3 Internal Connections 7 2 3 1 Opening the Housing 7 2 3 2 Jumper Connections 7 2 4 Levelling the Instrument 8 2 5 Wall Base Mount 8 2 6 Ini...

Page 4: ...set 19 3 7 2 Kilogram Unit 19 3 7 3 Gram Unit 19 3 7 4 Pound Unit 19 3 7 5 End Unit 19 3 8 Print Menu appears only with RS232 option installed 20 3 8 1 Reset 20 3 8 2 Baud 20 3 8 3 Parity 21 3 8 4 Sto...

Page 5: ...ERIAL COMMUNICATION 25 5 1 Interface Commands 25 5 2 Output Format 25 5 3 Printout Examples 26 6 MAINTENANCE 27 6 1 Cleaning 27 6 2 Troubleshooting 27 6 3 Service Information 27 7 TECHNICAL DATA 27 7...

Page 6: ......

Page 7: ...product Note For useful information about the product Warning Symbols General Hazard Electrical Shock Hazard 1 2 Safety Precautions For safe and dependable operation of this equipment please comply wi...

Page 8: ...view of Parts and Controls Item Description 1 Front Housing 2 Control Panel 3 Power Receptacle 4 Load Cell Cable Connector 5 RS232 Connector optional 6 Data Label 7 FCC Information 8 Rear Housing 9 Mo...

Page 9: ...ls Cont Main PC Board Item Description 1 Power Connector J1 2 Battery Connector J9 3 Option Connector J2 4 LFT Switch 5 4 6 Lines Sense Jumper W1 located on the other side of PCB 6 Load Cell Connector...

Page 10: ...t Item Description 1 Pound Kilogram Gram symbols 2 Battery function symbol 3 TARE Menu button 4 FUNCTION Mode button 5 Accumulation Symbol 6 NET function symbol 7 PRINT Units button 8 Dynamic symbol 9...

Page 11: ...ress Off Turns the Indicator on or off Units Changes the weighing Unit Mode Allows changing the application mode Press and hold allows scrolling through modes Menu Enter the User menu Menu Function Sh...

Page 12: ...or the power cord is removed Attention Before using the indicator for the first time the internal rechargeable battery should be fully charged for up to 12 hours The indicator can be operated during...

Page 13: ...ing the Housing CAUTION ELECTRICAL SHOCK HAZARD REMOVE ALL POWER CONNECTIONS TO THE INDICATOR BEFORE SERVICING OR MAKING INTERNAL CONNECTIONS THE HOUSING SHOULD ONLY BE OPENED BY AUTHORIZED AND QUALIF...

Page 14: ...l each time its location is changed 2 5 Wall Base Mount A Wall Base mounting bracket and screws are provided Base mounting Models C51XE6R C51XE30R C51XE50R C51XE200R Models C51XE30L C51XE50L C51XE100L...

Page 15: ...20 2d DYNAMIC GRAM 7 EVEN POWER ON UNIT 5d OFF OFF 7 ODD AUTO FILTER MANUAL ON 7 NONE GRAM LOW SEMI MAN LB 8 NONE KILOGRAM MED AUTO OFF STOP ZERO RANGE HI END MODE ON 1 0 AZT END UNIT 2 2 OFF HANDSHA...

Page 16: ...set to On 4 Allow the Indicator to warm up for approximately 5 minutes after stabilizing to room temperature 5 To abort calibration press the Exit button anytime during the calibration process 6 Make...

Page 17: ...ration Linearity calibration uses 3 calibration points The first calibration point is established with no weight on the scale The second calibration point is established at approximately half capacity...

Page 18: ...s in gravity Attention Changing the GEO Factor alters the calibration The GEO value was set at the factory and should only be changed by an authorized manufacturer s representative or certified verifi...

Page 19: ...14 13 13 12 12 11 11 10 40 05 41 12 16 15 15 14 14 13 13 12 12 11 11 41 12 42 19 16 16 15 15 14 14 13 13 12 12 11 42 19 43 26 17 16 16 15 15 14 14 13 13 12 12 43 26 44 32 17 17 16 16 15 15 14 14 13 1...

Page 20: ...ty Graduation Zero Range and settings are not reset 3 4 2 Capacity Set the scale capacity from 5 to 20000 Refer to the Setup Table 3 3 for available settings CAUTION Once the Capacity value has been c...

Page 21: ...CATAPULT 5000 SERIES SCALES EN 15 TABLE 3 3 SETUP AND CALIBRATION VALUES 6 15 0 0005 0 001 0 002 0 005 6...

Page 22: ...on Auto last unit in use when power was turned off kg g lb 3 4 5 Zero Range Set the percentage of scale capacity that may be zeroed 2 zero up to 2 percent of capacity 100 zero up to full capacity 3 4...

Page 23: ...drift of 1 division per second has been exceeded 3 d the display will maintain zero until a drift of 3 divisions per second has been exceeded NOTE When the LFT menu item is set to ON the selections a...

Page 24: ...resetting is initiated by pressing the FUNCTION key AUtO averaging is automatically initiated when the display is stable resetting is automatically initiated when the weight on the display is 1 divis...

Page 25: ...1 Reset Set the Unit menu to the factory defaults NO not reset YES reset 3 7 2 Kilogram Unit Set the status OFF Disabled ON Enabled 3 7 3 Gram Unit Set the status OFF Disabled ON Enabled 3 7 4 Pound U...

Page 26: ...nt 3 8 2 Baud Set the Baud rate 300 300 bps 600 600 bps 1200 1200 bps 2400 2400 bps 4800 4800 bps 9600 9600 bps 19200 19200 bps Reset No Yes Baud Rate 300 600 1200 2400 4800 9600 19200 Parity 7 Even 7...

Page 27: ...w control method NONE no handshaking ON OFF XON XOFF software handshaking 3 8 6 Stable Only Set the print criteria OFF values are printed immediately ON values are only printed when the stability crit...

Page 28: ...et weight is printed TARE OFF Tare weight is not printed ON Tare weight is printed UNIT OFF Unit is not printed ON Unit weight is printed MODE OFF Mode is not printed ON Mode is printed INFO OFF Info...

Page 29: ...peration 4 3 Manual Tare When weighing an item that must be held in a container taring stores the container weight in memory Place the empty container on the scale example 0 5 kg and press the TARE bu...

Page 30: ...accumulated data with no weight on pan press Function Mode key once The display blinks while showing the number of samples total weight Pressing the ZERO key zeros the display if required without aff...

Page 31: ...0T clears tare if allowed PU Print current unit g kg lb PCS xU Set scale to unit x 1 g 2 kg 3 lb xM Set scale to mode x 1M weigh 2M count 3M total 4M dynamic M will scroll to next enabled mode PV V Ve...

Page 32: ...Gross value line If Print Content Gross is ON and a tare value is entered 9 72 KG N Net value line If Print Content Net is ON and a tare value is entered 2 62 KG T Tare value line If Print Content Ta...

Page 33: ...y icon flashing Low Battery error Battery is empty CAL E Calibration Error Calibration value outside allowable limits Lo rEF Low reference weight warning message Average Piece Weight too small Warning...

Page 34: ...tainless steel Net Weight 4 kg 13 5 kg Shipping Weight 5 kg 15 kg Shipping Dimensions 400 x 365 x 210 mm 600 x 480 x 220 mm Scale Models C51XE100L C51XE200L C51XE30X C51XE50X C51XE100X C51XE200X Capac...

Page 35: ...T NUMBER RS232 Kit 30101019 In Use Cover 30101017 SF40A Impact Printer Contact OHAUS Lead Acid battery 30251812 7 3 Drawings and Dimensions Figure 8 1 C51 Overall Dimensions Dimention Chart mm A B C D...

Page 36: ...nt in EU Member States To comply with this Directive this device has been designed for safe removal of the batteries at end of life by a waste treatment facility Please dispose of this product in acco...

Page 37: ...t has been damaged by accident or misuse exposed to radioactive or corrosive materials has foreign material penetrating to the inside of the product or as a result of service or modification by other...

Page 38: ......

Page 39: ......

Page 40: ...ec des bureaux dans le monde entier Mit Niederlassungen weltweit Con uffici in tutto il mondo www ohaus com 30380009 P N 30380009B 2017 OHAUS Corporation all rights reserved todos los derechos reserva...

Reviews: