5
Capacity
65g
(2)
210g
(2)
150g
(1)
310g
(1)
410/100g
510g
1500g
3100g
4100/1000g
4100g
Readability
0.1mg
0.001g
0.01/0.001g
0.01g
0.1/0.01g 0.1g
Weighing units/modes
g, kg, mg, ct, N, lb, oz, ozt, gn, dwt, momme, mesghal, 3 taels, ticals, Parts Counting
Repeatability (Std. Dev.)
0.1mg
0.001g
0.01/0.001g
0.01g
0.1/0.01g
0.1g
Linearity
±0.2mg
±0.3mg
±0.002g
±0.01g/0.001
±0.01g
±0.02g
±0.1/0.01g
±0.1g
Tare Range
Full capacity by subtraction
Stabilization Time (seconds)
3 Seconds
Sensitivity Drift
4ppm/°C
10ppm/°C
Operating Temp Range
50° to 86° F/10° to 30° C
Power Requirements
8 - 14.5V 50/60Hz 6VA or 9.5 - 20V 6W
Calibration
External Digital
Display (in/cm)
LCD (1.2/3 high)
Pan Size (in/cm)
3.5/9 dia.
3.9/10 dia.
7.1/18 dia.
Draft shield clearance height
9.3/23.6
6.8/17.25
above pan (in/cm)
Dimensions (WxHxD) (in/cm)
8.5x14.3x13.5/
8.5x8.3x13.5/
8.5x4.3x13.5/21.7x11x34.3
36.6x11x34.3
21.7x21x34.3
Net Wt (lb/kg)
12.5/5.7
10/4.5
8.5/3.9
Shipping Wt (lb/kg)
21/9.5
16/7.3
14/6.4
Item No.
AR0640
AR2140 AR1530
AR3130
ARRV70
AR5120
ARA520 ARC120
ARRW60
ARD110
Error Codes
When a problem occurs using the balance, the display will indicated an error code. Review the listed codes and follow
instructions to correct the problem.
Err 2.0 Stability error. Check if balance is located near vibrating equipment or if air currents are affecting it.
Err 3.0 Calibration error (wrong mass). Incorrect or no calibration mass used when performing calibration procedure. Error
will flash momentarily, then balance will use previous calibration data. Recalibrate correctly.
Err 4
Invalid data checksum in type or adjust data. Return balance for servicing.
Err 6.0 Parts counting error - Average piece weight is less than 1d. Balance shows error then exits parts counting.
Err 8.4 Over or under load. Sample being weighed exceeds the capacity of the balance. If error occurs when the sample is
within the balance capacity, balance may be incorrectly calibrated. An underload such as the pan off of the balance
could also display Err 8.4. Recalibrate the balance.
Err 9
Internal data error. Return balance for servicing.
Err 9.8 Invalid checksum in calibration or setup data. The balance may need recalibration, particularly linearity calibration.
If the error persists after recalibration, thebalance must be serviced.
SPECIFICATIONS
Installation Category II, Polution Degree 2.
Note 1. 150 g and 310 g balances include a round draft shield.
Note 2. 65 g and 210 g analytical balances include a three door draft shield.