4

CALIBRATION MASSES

                                       LINEARITY             SPAN ONLY
           CAPACITY              MASSES                MASSES

65g

20g/50g

50g

150g

100g/150g

150g

210g

100g/200g

200g

310g

100g/300g

300g

 510g

300g/500g

500g

1500g

500g/1500g

1500g

3100g

1000g/3000g

 3000g

4100g

2000g/4000g

4000g

CALIBRATION (Cont.)

CARE AND MAINTENANCE

To keep the balance operating properly, keep the housing
and platform clean. If necessary, a cloth dampened with a
mild detergent may be used. Keep calibration masses in a
safe dry place.

RS232 COMMAND DATA TABLE

Command
Character

Description

P

Print command

T

Tare command

?

Displays current mode

XS

X=0  (zero) Stable off

XS

X=S Stable on (default setting in balance)

XS

X=A Auto print on stabilty

Masses must meet or exceed ASTM Class 1 Tolerance.
Calibration masses are available as accessories.

Calibration Masses

Before beginning calibration, make sure masses are available.
If you begin calibration and realize calibration masses are
not available, exit the procedure by pressing and holding Mode
Off
 until balance returns to weigh mode. The balance will
retain previously stored calibration data. Calibration should
be performed as necessary to ensure accurate weighing.
Masses required to perform the procedures are listed in the
following table.

Span Calibration

1. With the balance turned ON, press and hold >O/T< until

CAL is displayed.

2. Release >O/T<-C- momentarily displayed followed by the

value of the calibration mass which is to be placed on the
pan.

Do not disturb the balance when -C- is displayed.
Incorrect calibration may result.

3. Place the indicated span calibration mass on the pan.

4. Press >O/T<, -C- is momentarily displayed, then the weight

of the mass on the pan is displayed.

5. Remove the calibration mass from the pan. The balance is

 now calibrated and returns to the weighing mode.

Linearity Calibration

1. With the balance OFF, press and hold >O/T< until MENU

is displayed, then release it and UNITS is displayed. Press
Mode Off, LIN is displayed.

2. Press >O/T<-C- is displayed followed by the value of the

first mass which must be placed on the pan.

3. Place the required mass on the pan  and momentarily press

>O/T<, -C- is displayed followed by the value of the next
 mass to be placed on the pan.

Do not disturb the balance while -C- is displayed.

4. Place the second required mass on the pan and

momentarily press >O/T<, -C- is displayed. When the
weight on the pan is displayed with the stability indicator,
the balance is calibrated and returns to the weighing mode.

5. Remove the calibration masses from the pan.

1

N/C

2

Data Out (TXD)

3

Data In (RXD)

4

N/C

5

Connected to pin 8

6

N/C

7

Ground

8

Connected to pin 5

9

N/C

SYMPTOM

No Display.

Incorrect weight
reading.

Calibration procedure
does not work.

Unable to display
weight in a particular
weighing unit.

Balance won't store
selections made in
menu.

PROBABLE CAUSE

Power Adapter not
connected.

Balance out of
calibration.

Balance was not
rezeroed before
weighing.

Incorrect calibration
masses being used.

Weighing unit not
activated in menu.

END selection was not
used to exit menu.

REMEDY

Connect AC Adapter.

Calibrate the balance.

Press >O/T<  with no
weight on the pan,
then weigh item.

Use correct masses.

Use Units menu to set
desired units ON (see
menu).

You must use END to
exit each menu and
save selections.

TROUBLESHOOTING

NOTE: Print commands entered through the computer are
temporary. When the balance is turned off, it will return to
balance menu settings when turned on again.

1

2

3

4

5

6

7

8

9

Summary of Contents for Adventurer series

Page 1: ...i INSTRUCTION MANUAL AdventurerTM Balances M ode Off On On Print Mode Off On On Print...

Page 2: ...EC EN55011 1991 class B Radiointerferenze Compatibilit elettromagnetica EN50082 2 1995 Immunit ISO 9001 Registration for Ohaus Corporation Ohaus Corporation USA was examined and evaluated in 1994 by t...

Page 3: ...UNG VERWENDET WIRD KANN DER FUNKVERKEHR GEST RT WERDEN DURCH DEN BETRIEB DIESES GER TS IN EINEM WOHNGEBIET K NNEN DERARTIGE ST RUNGEN VERURSACHT WERDEN IST DIES DER FALL MUSS DER BENUTZER DIESE ST RUN...

Page 4: ...anel All AdventurerTM balances are initially factory set to measure in grams and can be set to measure in kilograms milligrams carats Newtons pounds ounces troy ounces grains pen nyweight mommes mesgh...

Page 5: ...setup procedure 2 If it is necessary to rezero the display momentarily press O T 3 Place item s to be weighed on the pan and read the weight on the display The stability indicator appears when the re...

Page 6: ...utomatic stable read ings is OFF Only one of these may be set on at a time Stable Data On or Off When STBL ON is displayed press Mode Off to select ON or OFF then press O T AUTO OFF is displayed Auto...

Page 7: ...then the weight of the mass on the pan is displayed 5 Remove the calibration mass from the pan The balance is now calibrated and returns to the weighing mode Linearity Calibration 1 With the balance O...

Page 8: ...listed codes and follow instructions to correct the problem Err 2 0 Stability error Check if balance is located near vibrating equipment or if air currents are affecting it Err 3 0 Calibration error...

Page 9: ...oston Road Boite Postale 30 An der Johanneskirche 6 Motores 178 Beaumont Leys 85 rue Joseph Bertrand D 35390 Giessen 08038 Barcelona Leicester LE4 1AW 78220 Viroflay Deutschland Espana England France...

Reviews: