CHAPTER 2 TROUBLESHOOTING
Ohaus Corporation www.ohaus.com
2-14
7000 Series T72XW Indicators Service Manual
Figure 2-8: Switches SW1-2 and SW1-4
The position of SW1-4 (also shown in
Figure
) determines whether metrologically significant EEPROM data for scale calibration is reset
when a master reset is performed. SW1-4 must be set to ON to reset EEPROM data. If SW1-4 is set
to OFF, metrological EEPROM data will not be affected by the master reset and the original
calibration will persist.
5
Apply AC power. During the power up sequence, the display will indicate a warning message of
[WARNING MASTER RESET! Press
ENTER
to continue.
6
Press
ENTER
to perform a master reset. When the reset is complete, the screen will return to the
normal weight display.
OR
To exit without a master reset, do not press
ENTER
.
7
Remove AC power.
8
Return SW1-2 (and, if it was changed, SW1-4) to their original OFF positions.
9
Reapply AC power.
2.13
Upgrading Firmware
Contact your local authorized OHAUS CORPORTATION service representative for information
on upgrading firmware.