Ofs Softening Control Instruction Manual Download Page 20

 
 

 

 

 

 

20 

 

 

10.3

 

Operating tests 

 

 
10.3.1

 

Test of water quality  

 
By temporary interruption of power supply Softening Control  re-starts the measurement and 
proceeds a new calibration independent of preset measuring cycle. Provided that the quality of 
soft  water  is  really  below  the  parameterized  threshold  value,  upon  completing  the  measure-
ment result will be displayed by the luminescent green LED and the current loop comes into 
operation. 
 
In case of exceeding the threshold value re-calibration will be done. If this results once again 
in an exceeding threshold value, an external threshold value alarm will be signalized accord-
ing to the preset alarm delay at DIP 6 and the measurement will be repeated until a lower de-
viation is discoverable cyclically. 
 

 
 
10.3.2

 

Test of external signaling 

 

A  threshold  value  can  be  initiated,  if  the  soft  water  supply  will  be  external  and  manually 
closed,  whereby  the  raw  water  cannot  be  flushed  out  of  the  measuring  chamber  by  the  soft 
water. With repeated recognition of the exceeded threshold value the relevant threshold value 
relay will be switched after expiration of the delay preset at DIP 6 until a lower deviation is 
discoverable again. 
 
A  service  alarm  can  be  initiated  by  disconnection  of  the  plug-in  connection  to  the  hardness 
sensor.  Upon  a  short  time  the  green  and  red  LED  will  be  flashing  and  the  relay  for  service 
alarm will be switched until the sensor will be recognized. 
 
 

10.3.3

 

Test sensor performance 

 
If, after short operating time of the sensor, sensor alarm is output (all three LED are flashing), 
the sensor quality can checked once more by a further calibration. The calibration can be trig-
gered by a short-term interruption of the power supply. If the sensor alarm appears again, the 
sensor must be replaced. 
 
If the operating time of the sensor is significantly below the indicated operating time, the sen-
sor should be sent to the manufacturer for review. 
 

Summary of Contents for Softening Control

Page 1: ...Softening Control Instruction Manual Version 1 13...

Page 2: ......

Page 3: ...ter 14 9 3 6 1 Connection mechanical water meter 14 9 3 6 2 Connection electronic water meter turbine with Hall contacts 14 9 3 7 Overview electrical connections 15 10 Commissioning 16 10 1 Setting of...

Page 4: ...4...

Page 5: ...ce incorrect or unauthorized intervention of the device or its components incorrect analytical mechanical or electrical effects of any kind damages or subsequent damages due to incorrect operating dam...

Page 6: ...safe and hazard free operation of the device Caution With disregarding of so marked information the device or sensor could be damaged Warning The disregarding of so marked information may cause injur...

Page 7: ...tion must not exceed the specified voltage The raw water and soft water of the monitored or controlled process must cor respond to the relevant drinking water specifications During operation a constan...

Page 8: ...er and will be evaluated after a preset resting time The measurement cycle can be controlled alternatively by impulses of a connected water meter or a potential free closer By automatic opening and cl...

Page 9: ...value given by a DIP switch The difference value shows the actual efficiency of the softening plant The threshold value can be defined individually in steps of 0 25 up to 10 in refer ence to the row...

Page 10: ...an be installed easily 9 2 Hydraulic installation The hydraulic installation of inlets outlet and hardness sensor is only to be carried out after the unit has been fixed at the wall resp frame as foll...

Page 11: ...h would otherwise reach the sampling valves optional connection water meter optional external signal output power supply unit 100 240 V AC 50 60 Hz reserve connection measurement chamber Ventil RW Ven...

Page 12: ...3 2 Output water hardness by active current loop The signal is a coded residual hardness signal and the current of 4 20 mA represents a re sidual water hardness in the range of 0 2 or 0 10 The curren...

Page 13: ...ED status for displaying of functional status and Malfunctions and possible causes The following external electrical requirements must absolutely be fulfilled External requirements maximum connection...

Page 14: ...ng alarm caused of the so called hardness breakthrough it is recommend to monitor the soft water sampling not within fixed time distances Therefore it is possible to connect a mechanical or electrical...

Page 15: ...oop resistor 250 Ohm 18 4 20 mA Loop resistor 250 Ohm 20 Offline contact Potential free contact 21 Offline contact Potential free contact 23 Water meter or potential free closer Potential free contact...

Page 16: ...d Accord ing this decision a connection of contact wa ter meter and setting of DIP switch 4 is to realize The threshold value to be monitored is to set up with the DIP switches 1 3 The following proce...

Page 17: ...1 DIP 2 DIP 3 Threshold value Example With a raw water hardness of 5 0 mmol l alarm will be signalized at ON ON ON 0 25 0 013 mmol l OFF ON ON 0 50 0 025 mmol l OFF ON OFF 1 00 0 05 mmol l OFF OFF ON...

Page 18: ...evaluated accordingly LED Display Evaluation Example green LED lights up water hardness 80 threshold value threshold value 1 water hardness 0 8 green and yellow LED are lighting up water hardness 90 t...

Page 19: ...ves for tightness clean or change valves green and red LED is flashing sensor not recognized Service alarm The measuring system has no elec trical contact to the hardness sen sor check the plug in con...

Page 20: ...e external and manually closed whereby the raw water cannot be flushed out of the measuring chamber by the soft water With repeated recognition of the exceeded threshold value the relevant threshold v...

Page 21: ...sed and with internal humidity a delivery without prejudice is realized A comparison of the number on the sensor with the numbers indicated at packaging and delivery note is rec ommended Further the s...

Page 22: ...sor is mounted and remove the sensor straight from the flow measuring chamber 4 Take the new sensor out of the box and pull it out of the protective cap straightly Do not touch or damage the sensible...

Page 23: ...sensor depending on concentration and duration A strong deep black colouring of the metal electrodes indicates high chlorine influence The sensor may indicate threshold value exceeding while no thresh...

Page 24: ...LED are flashing sensor cable is not properly plugged in or electrical plug in connection defective examine the electrical connection between sensor and device all LED are flashing Sensor performance...

Page 25: ...nsor Protect the sensible sensor part by a protective cap filled with approx 1 ml raw water Store the sensor in a cool 10 15 C and dark place 14 Dismounting 1 Take the device out of operating as descr...

Page 26: ...ter 0 18 mmol l up to 10 8 mmol l natural water Evaluation of water hardness in raw water 100 Evaluation of water hardness in soft water parameterizable threshold value between 0 25 and 10 in referenc...

Page 27: ...rangezogen For verification in accordance with Low Voltage Directive 2014 35 EU the following standards were applied EN 61010 1 2010 Zum Nachweis entsprechend EMV Richtlinie 2014 30 EG wurden folgende...

Page 28: ...OFS Online Fluid Sensoric GmbH Heidelbergweg 9 07580 Ronneburg Germany Manual Softening Control Version 1 13 Issue 12 2018 Subject to technical changes...

Reviews: