4. Place the cell in the heating jacket with the outlet or filter side of the cell
cap properly oriented down. Make sure both valves on the cell are
closed. Transfer the thermometer into the thermometer well within the
cell.
5. Check the o-ring on the inside of the female coupling on the top pres-
sure assembly for nicks and cuts. Apply a thin coat of silicone grease
to the o-ring but do not seal up the hole with grease. Connect the pres-
suring assembly to the top male slip coupling by lifting the lock ring and
sliding the pressure assembly onto the male coupling. Release the lock
ring and the top pressure assembly has been installed.