background image

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com

6

Calibration - 

Performance 

Test

1.  Inspect the electrode probe and cable for evidence of damage.  Make 

sure the entire electrode gap is free of deposits and the connector to the 

instrument is clean and dry.

2.  Disconnect the electrode probe from the meter and push the button to 

run a voltage ramp test.  If the meter is working properly, the ES reading 

should reach the maximum voltage permitted by the instrument.

3.  Plug the high calibration standard into the same port the probe was in.  

Run the voltage ramp test.  The ES value should be within 2% of the val-

ue marked on the standard.  Repeat this process with the low calibration 

standard.  Again, the ES value should be within 2% of the value marked 

on the standard.  

If either reading is not within the specified range, the unit will require ser

-

vice from a trained technician.

Calibration Standards

4.  Reconnect the electrode to the meter and repeat the voltage ramp test 

using tap water.  The ES reading should not exceed 6 volts.  If the reading 

does exceed 6 volts, clean the electrode probe again using a solvent (Iso

-

propanol, etc.) and carefully check the electrical connector for damage or 

wear.  If the unit is still not functioning properly, the electrode and cable 

may be replaced by calling OFITE and ordering a new electrode (part no. 

131-01).

 

These procedures do not check the waveform, AC frequency or ramp 

rate. The ES meter must be periodically checked by a qualified technician 

to ensure that all units meet specifications.  OFITE recommends returning 

the ES Meter once a year for a thorough calibration check.

Summary of Contents for 131-50

Page 1: ...lity Meter Part No 131 50 Instruction Manual Updated 12 10 2015 Ver 3 0 OFI Testing Equipment Inc 11302 Steeplecrest Dr Houston Texas 77065 U S A Tele 832 320 7300 Fax 713 880 9886 www ofite com Copyr...

Page 2: ...rest Dr Houston TX 77065 USA Tel 832 320 7300 Fax 713 880 9886 www ofite com 1 Intro 2 Description 2 Components 3 Specifications 4 Procedure 5 Calibration Performance Test 6 Warranty and Return Policy...

Page 3: ...age is the point at which the drilling fluid be comes electrically conductive The DC power source of the batteries pro vides an AC voltage to the electrodes at a low frequency The unit provides a cons...

Page 4: ...tion Standard 147 02 9 Volt Alkaline Battery Optional 110 10 Marsh Funnel Viscometer or 12 mesh screen 130 38 Thermocup 115 Volt 130 38 1 Thermocup 230 Volt 154 01 Dual Scale Thermometer with Metal Di...

Page 5: ...3 2 000 Volts Peak Trip Current 61 5 mA Size 9 5 6 5 3 5 24 1 16 5 8 9 cm Weight 2 lbs 15 oz 1 3 kg Shipping Size 11 11 7 28 28 18 cm Shipping Weight 10 lbs 4 54 kg Electrode Housing Material resistan...

Page 6: ...120 F 48 8 C 3 Pour the sample through a 12 mesh screen or Marsh Funnel screen into a glass or plastic container Do not use a metal container Record the mud temperature on the Drilling Mud Report Form...

Page 7: ...andard Again the ES value should be within 2 of the value marked on the standard If either reading is not within the specified range the unit will require ser vice from a trained technician Calibratio...

Page 8: ...ny losses or damages that occur as a result of Improper installation or maintenance of the products Misuse Neglect Adjustment by non authorized sources Improper environment Excessive or inadequate hea...

Reviews: