OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
11
Setup
Temperature and
Pressure Controller
Page Scroll
The Eurotherm Model 3504 controls both temperature and pressure. When
running a test with the computer, the software automatically programs the
controller. However, when running a test without the computer, the controller
must be programmed manually. It is strongly recommended that operators
carefully study the Model 3504 instruction manual before attempting to
program the controller.
When running a test without the computer, always build the program prior to
mixing the slurry and loading the slurry cup.
The Eurotherm 3504 can store up to eight custom programs. Each program
consists of multiple segments. In a segment, the temperature or pressure
setpoint can be changed or held for a certain amount of time.
Each program also consists of two channels. Channel 1 controls the
temperature and Channel 2 controls the pressure. The two channels are
programmed separately and have their own set of segments.
Selector
Temperature
Pressure
To edit a program:
1. Press the “Page” button until the display reads “Program Edit”.
2. Use the up and down arrows to select the program you want to edit.
3. Press the “Scroll” button to move to the Channel parameter.
4. Use the up and down arrows to select the Channel you want to edit.
Channel 1 controls temperature. Channel 2 controls pressure.