background image

- 9 -

994293-Z00

PUTTING INTO OPERATION AND SETTING

UVEDENÍ DO PROVOZU A NASTAVENÍ

9

Several basic parameters need to be set to put the PAQ into operation. Follow the instructions below:

1. Connect the device according to the selected diagram.

2. Connect the correct voltage according to the nameplate on the back of the device.

3. Press the SET key for at least 5 seconds to enter the configuration mode.

4. Enter the P_1 menu by pressing the SET key.

5. If a voltage transformer is used, set its ratio in the parameter Utr. The ▲ key is used to switch parameters 

within the menu. The SET key activates editing and saves the newly set value. The value can be changed 

using the cursor keys ▲ (+) and ▼ (-).

6. Set the ratio of the measuring current transformer in the parameter Itr. Use the ▲ (+) and ▼ (-) keys to change 

the value. Confirm the new ratio by pressing the SET key.

7. Check the device frequency used, parameter Fr (50 or 60 Hz).

8. To exit the P_1 menu press the SET key. Pressing the ESC key again exits the configuration mode and 

returns the device to normal monitoring mode.

9. Check the measured values for correctness.

Pro uvedení analyzátoru PAQ do provozu je zapotřebí nastavit několik základních parametrů. Postupujte podle 

následujících instrukcí:

1. Zapojte přístroj podle vybraného schématu.

2. Připojte správné napájecí napětí dle štítku na zadní straně přístroje.

3. Stiskněte klávesu 

SET

 po dobu nejméně 5 sekund pro vstup do konfiguračního režimu.

4. Vstupte do menu 

P_1

 stisknutím klávesy 

SET

.

5. Pokud je použit měřicí transformátor napětí, nastavte jeho převodový poměr v parametru 

Utr

. Klávesa ▲ je použita 

pro přepínání parametrů v rámci menu. Klávesa 

SET

 aktivuje editaci a ukládá nově nastavenou hodnotu. Změna 

hodnoty je možná pomocí kurzorových kláves ▲ (+) a ▼ (-).

6. Nastavte převodový poměr měřicího transformátoru proudu v parametru 

Itr

. Pro změnu hodnoty použijte klávesy ▲ 

(+) a ▼ (-). Nově nastavený převodový poměr potvrďte klávesou 

SET

.

7. Zkontrolujte použitou frekvenci přístroje, parametr 

Fr

 (50 nebo 60 Hz).

8. Pro návrat z menu 

P_1

 stiskněte klávesu 

ESC

. Další stisknutí klávesy 

ESC

 ukončí konfigurační režim a vrátí přístroj 

do normálního monitorovacího režimu.

9. Zkontrolujte měřené hodnoty zda odpovídají skutečnosti.

The PAQ settings are divided into three separate menus in the configuration 

mode.

Press the SET key for at least 5 seconds to enter the configuration mode. The 

following screen appears on the device's display.

Use the cursor keys ▲ and ▼ to move through the menu. The ▲ key is normally 

used to move around within the selected menu. Parameter editing is activated by 

the SET key and the value of the parameter is changed by the cursor keys ▲ and 

▼.

The newly set value is confirmed and saved by pressing the SET key.Text EN

The ESC key can be used to terminate the parameter editing process, return to 

the higher menu or exit the configuration mode at any time.

Nastavení analyzátoru PAQ je rozděleno do tří samostatných menu v konfiguračním 

režimu.

Pro vstup do konfiguračního režimu stiskněte klávesu SET po dobu nejméně 5 sekund. 

Na displeji přístroje se zobrazí následující obrazovka.

Pro pohyb v menu slouží kurzorové klávesy ▲ a ▼. Klávesa ▲ je normálně použita pro 

pohyb dokola v rámci zvoleného menu. Editace parametru je aktivována klávesou SET 

a změna hodnoty daného parametru kurzorovými klávesami ▲ a ▼.

Nově nastavená hodnota je potvrzena a uložena po stisku klávesy 

SET

.

Klávesou 

ESC

 je možné kdykoliv ukončit proces editace parametru, provést návrat do 

nadřazeného menu či ukončit konfigurační režim.

Prog

Summary of Contents for PAQ-50 Series

Page 1: ...NALYZER ANALYZ TOR PAQ 50 PAQ 51 PAQ 52 INSTRUCTIONS FOR USE N VOD K POU IT Installation service and maintenance of the electrical equipment may be carried out by an authorized person only Mont obsluh...

Page 2: ...ement using current transformers X 5 or X 1 The multifunction network analyzer is designed for monitoring electrical parameters of three phase and single phase LV and MV networks The device design is...

Page 3: ...99 0 999 5 TDD Total demand distortion TDD 0 999 5 Voltage harmonics Pod l vy ch harmonick ch U 0 999 0 999 class 1 Current harmonics Pod l vy ch harmonick ch I 0 999 0 999 class 1 Voltage asymmetry N...

Page 4: ...ion in the menu value increase Pohyb v menu zv en hodnoty ESC to cancel selection exit the menu ESC zru en volby odchod z menu Navigation in the menu value decrease Pohyb v menu sn en hodnoty Active s...

Page 5: ...5 6 7 8 9 10 l k l k l k I1 I2 I3 1 2 3 4 L1 L2 L3 N 16 15 14 13 B RS485 A 24 23 22 21 20 19 R3 R2 R1 25 26 27 28 29 30 COM IN IO4 IO3 IO2 IO1 COM OUT 12 L 11 24VDC AC N 5 6 7 8 9 10 l k l k l k I1 I...

Page 6: ...6 994293 Z00 2 2 4 5 1 3 MOUNTING MONT 6 dimensions rozm ry 88 54 42 6 27 3 5 45 2 62 10 90 2 109 5 88 10 54 42 6 17 5 27 3 5 45 2 62 90 2 109 5 PAQ 50 PAQ 51 PAQ 52...

Page 7: ...ment 0 HQt QDS Wt Current measurement 0 HQt SURXG PAQ 5 U230 PAQ 5 U024 Power supply 1DSiMHQt L1 L2 L3 N k l k l k l N FU F1 FU L1 L2 L3 N PE L K L K L K L k l k l k l 11 12 Power supply 1DSiMHQt N L...

Page 8: ...another device such as an RS485 signal converter to another interface or a master device equipped with a communication gateway P stroj je vybaven galvanicky odd len m komunika n m rozhran m RS485 s p...

Page 9: ...T aktivuje editaci a ukl d nov nastavenou hodnotu Zm na hodnoty je mo n pomoc kurzorov ch kl ves a 6 Nastavte p evodov pom r m ic ho transform toru proudu v parametru Itr Pro zm nu hodnoty pou ijte kl...

Page 10: ...Y year 20 nastaven roku 20 00 09 99 M month setting nastaven m s ce 01 01 12 D day setting nastaven dne 01 01 31 h hour setting nastaven hodiny 00 00 23 M minute setting nastaven minuty 00 00 59 PAS p...

Page 11: ...11 994293 Z00 Prog 5s Prog Prog Prog Prog Prog kV Prog kV Prog kVA Prog kVA Prog Hz kWh Prog kWh Prog Prog Hz...

Page 12: ...12 994293 Z00 Prog Prog Prog Prog Prog Prog Prog Prog Prog Prog Prog Prog Prog Prog Prog...

Page 13: ...ozsah nastaven Id unique device number in the RS485 network jedin n identifika n slo p stroje v s ti RS485 0 0 255 bd communication speed of data transmission komunika n rychlost p enosu dat 9 6 9 6 1...

Page 14: ...irmware maz n energi Parameter Parametr Meaning V znam FIr firmware version information informace o verzi firmware VEr hardware version verze hardware CLr clearing saved tariff energy nulov n po tadla...

Page 15: ...COS phi COS fi Active power LQQ Y NRQ Reactive power Jalov v kon SSDUHQW SRZHU GiQOLY Y NRQ The device is equipped with clear user display with symbols of displayed values for measured quantities For...

Page 16: ...a tes with the meanings described in the following table Signalizace stavu vstup a v stup Vstupy a v stupy se mohou nach zet ve ty ech provozn ch stavech Signalizace na displeji je spole n pro v echny...

Page 17: ...z kladn nastaven p mo pomoc tla tek na p stroji Dal funkce jsou dostupn pouze ze softwaru OEZ PAQ Lze tak p ehledn a jednodu e konfigurovat nastaven nastavovat p esn as definovat parametry komunikace...

Page 18: ...and time zone as a datum Nastaven asu a data z PC a mo nost nastaven p echodu letn ho zimn ho asu a asov z ny RS485 communication Setting RS485 ID baud rate parity stopbit Komunikace RS485 Nastaven I...

Page 19: ...similar to PLC programming Individual logic operations are defined for different function blocks In this universal way it is possible to define functions of e g current relay or more complex three sta...

Page 20: ...stored Consideration should be given to the appropriate storage interval given the potentially large amount of data in the database Z znam hodnot OEZ PAQ umo uje zaznamen vat vybran m en hodnoty v dan...

Page 21: ...s Nastaven tarifu Nastaven ovl d n tarifu pro m en energi Tarif m e b t p ep n n na z klad I O vstupu nebo pomoc asov ch program Modbus registers A detailed description of the Modbus registers is avai...

Page 22: ...stup vstup 3 relays 5 A 250V AC 4 digital I O Type of digital outputs Typ digit ln ch v stup open collector optical separation otev en kolektor optick odd len Max voltage of digital outputs Max nap t...

Page 23: ...pom r m c ho transform toru nap t 1 750 000 Ratio of measuring current transformer P evodov pom r m c ho transform toru proudu 1 10 000 Max number of registered failures Max po et registrovan ch v pa...

Reviews: