COMPACT 7600 Periodic Maintenance, 00-878852-01

4COMPACT 7600 Pe

4.6 Perform System Functional Test

WARNING:

Steps within this procedure produces X-rays. Use appropriate precautions.

NOTE:

Use the illustration below as a reference when performing the Operational Verification.

1.

Press the C-up and C-down buttons 

 and verify that the c-arm lifts smoothly from

top to bottom limits.

2. Press the Auto Fluoro button 

 repeatedly and verify that the LED turns on and off. With

Auto Fluoro active (LED on) press the x-ray button and verify that the kV and mA automatically
servo by monitoring the kV and mA display. Verify that an image appears on the monitor. Verify
that the mA display shows the mA value while x-rays are being taken.

3. Press the Pulse Fluoro button 

 repeatedly and verify that the LED turns on and off. With

Pulse Fluoro active (LED on) press the x-ray button and verify that the system produces pulsed
x-rays and that an image appears on the monitor.

4. Press the Snapshot button 

 repeatedly and verify that the LED turns on and off. With

Snapshot active (LED on) press the x-ray button and verify that an exposure is taken and then
automatically terminates. The image should remain on the monitor until the next exposure is
taken.

5. Press the Auto Fluoro button 

 and then the kV button 

 and verify that both LED’s

illuminate. Press the kV-up/down button and verify that the kV can be changed from 36 to
110kV.

6. Press the Film button 

 repeatedly and verify that LED turns on and off. With Film active

(LED on) press the Film Timer-up/down button and verify that you can select a time from 0.1
sec. to 4 sec. as shown on the seconds display. Take a number of film shots at various kV and
mAs settings to verify film mode operation.

Used and Refurbished C-Arms are Available from www.SharpMedical.com - Call us at 800-969-9800 

An Independent C-Arm Service Provider.  This PDF provided for research / historical purposes only.

Summary of Contents for COMPACT 7600

Page 1: ...Maintenance Procedure 00 878852 01 Used and Refurbished C Arms are Available from www SharpMedical com Call us at 800 969 9800 An Independent C Arm Service Provider This PDF provided for research his...

Page 2: ...ne who is not specifically trained or authorized by OEC Medical Systems Inc to work on the COMPACT 7600 System The contents of this procedure are believed accurate at the time of publication However c...

Page 3: ...996 Initial Engineering Release Rev 2 01 Aug 1996 Added fast stop button operational check Rev A 01 Aug 1996 Manufacturing Release Used and Refurbished C Arms are Available from www SharpMedical com C...

Page 4: ...riodic Maintenance 00 878852 01 Used and Refurbished C Arms are Available from www SharpMedical com Call us at 800 969 9800 An Independent C Arm Service Provider This PDF provided for research histori...

Page 5: ...4 5 Measure Line Voltage Regulation 3 4 6 Perform System Functional Test 4 5 Generator Accuracy Verification Calibration 6 5 1 Verify X Ray Calibration 6 5 2 Perform X Ray Calibration 6 6 Radiographic...

Page 6: ...riodic Maintenance 00 878852 01 Used and Refurbished C Arms are Available from www SharpMedical com Call us at 800 969 9800 An Independent C Arm Service Provider This PDF provided for research histori...

Page 7: ...Serious injury and property damage can result from incorrectly performed service procedures Observe all operating and safety procedures contained within this manual CAUTION The procedures in this sec...

Page 8: ...1 Mechanical Checks In the steps below check for ease of movement without excessive play 1 Check the Locking Pedals of the system for proper operation Lock the system in place by pressing the locking...

Page 9: ...and has a 9 pin connector on the end of the bundle The 9 pin connector is designated as P2 Measure the voltages listed below and record them on the PM Form GND To Adjust Voltage Tolerance P2 2 P2 1 R3...

Page 10: ...button and verify that the system produces pulsed x rays and that an image appears on the monitor 4 Press the Snapshot button repeatedly and verify that the LED turns on and off With Snapshot active...

Page 11: ...the image rotates to the normal orientation and automatically stops 13 Press the Thorax Surgery button repeatedly and verify that the LED turns on and off With Thorax Surgery active LED on take an x r...

Page 12: ...n of the Generator Accuracy Verification fails to meet specifications proceed to section 5 2 Perform X Ray Calibration of the PM procedure TEST TECHNIQUE KV TOLERANCE MA TOLERANCE Manual Fluoro 60kV 5...

Page 13: ...section are designed to ensure compliance with the 10 0 R minute entrance exposure limitation set by the Bureau of Radiological Health BRH Refer to the MA CALIBRATION section of the COMPACT 7600 Servi...

Page 14: ...ntenance 00 878852 01 8COMPACT 7600 Pe Used and Refurbished C Arms are Available from www SharpMedical com Call us at 800 969 9800 An Independent C Arm Service Provider This PDF provided for research...

Reviews: