COMPACT 7600 Periodic Maintenance, 00-878852-01
4COMPACT 7600 Pe
4.6 Perform System Functional Test
WARNING:
Steps within this procedure produces X-rays. Use appropriate precautions.
NOTE:
Use the illustration below as a reference when performing the Operational Verification.
1.
Press the C-up and C-down buttons
and verify that the c-arm lifts smoothly from
top to bottom limits.
2. Press the Auto Fluoro button
repeatedly and verify that the LED turns on and off. With
Auto Fluoro active (LED on) press the x-ray button and verify that the kV and mA automatically
servo by monitoring the kV and mA display. Verify that an image appears on the monitor. Verify
that the mA display shows the mA value while x-rays are being taken.
3. Press the Pulse Fluoro button
repeatedly and verify that the LED turns on and off. With
Pulse Fluoro active (LED on) press the x-ray button and verify that the system produces pulsed
x-rays and that an image appears on the monitor.
4. Press the Snapshot button
repeatedly and verify that the LED turns on and off. With
Snapshot active (LED on) press the x-ray button and verify that an exposure is taken and then
automatically terminates. The image should remain on the monitor until the next exposure is
taken.
5. Press the Auto Fluoro button
and then the kV button
and verify that both LED’s
illuminate. Press the kV-up/down button and verify that the kV can be changed from 36 to
110kV.
6. Press the Film button
repeatedly and verify that LED turns on and off. With Film active
(LED on) press the Film Timer-up/down button and verify that you can select a time from 0.1
sec. to 4 sec. as shown on the seconds display. Take a number of film shots at various kV and
mAs settings to verify film mode operation.