![OEC 9600 Maintenance Procedure Download Page 35](http://html1.mh-extra.com/html/oec/9600/9600_maintenance-procedure_3377918035.webp)
9600 Periodic Maintenance Procedure
00-880606-02
Page 35
2. Reach through the access holes in the casting and spray the horizontal and vertical chains lightly
with Permatex Super Lube Spray P/N - 88-299485-00.
5. X-RAY BEAM ALIGNMENT VERIFIED (annual pm only)
WARNING
The following steps produce X-rays. Use appropriate precautions.
Setup
1. To nominalize C-Arm flexure, release the C-Arm brake and position the C-Arm at a 45
°
angle as
shown below.
2. Attach the Beam Alignment Tool on the Image Intensifier.
C-Arm Positioning
45°