9600 Periodic Maintenance Procedure
00-880606-02
Page 23
BOOST/ENABLE*
1. Press
the
BOOST/ENABLE
key for approx. 5 seconds and verify BOOST is selected and its LED
lights.
*Cardiac systems are always in the boost mode.
IMAGE REVERSAL
1. While taking X-rays, press the
IMAGE REVERSAL
key and verify image reversal for each of four
image positions. The image position’s LED should light when selected.
KVP CONTROL
1. Select Manual Fluoro Mode, then press the
KVP CONTROL
key and verify kVp can be adjusted.
MA CONTROL
1. Press
the
MA CONTROL
key and verify that mA can be adjusted in
MANUAL FLUORO
mode
and mAs can be adjusted in
FILM
mode.
ABS SELECT
1. Press
the
ABS SELECT
key and verify the different ABS tables can be selected and loaded.
2. POWER SUPPLIES & BATTERY INSPECTIONS
2.1 SYSTEM BATTERIES
GENERATOR BATTERIES
This procedure can be used to verify that a fully charged set of battery packs is performing adequately.
WARNING
The following steps produce X-rays. Use appropriate precautions.
1. Remove the Torx screw on the right side of the electronic Card Rack and swing the Card Rack out
on its hinge to expose the Battery Charger PCB.
2. In
FILM
mode, set the technique for 75 kVp @ 200 mAs.
3. Make an exposure and verify that the Bargraph LED located on the Battery Charger PCB indicates
160 or 170 Volts.
4. If the bargraph voltage indicates less than 160 Volts, measure across the battery terminal with a
DMM while making an exposure. If the batteries measure less than 160 Volts, weak battery packs or
a fault on the Battery Charger PCB may be indicated.