4 Safety In
s
truction
s
Instruction Manual Corvis® ST (G/72100/XXXX/EN 1019 Rev01)
9 / 80
Ensure that connections with non-medical devices are made
correctly.
Only use the power adapter listed in the packing list.
Establish an USB connection only with the OCULUS USB FS
MEDIsolator (Nr. 01 56920 00 010).
Use only a computer that meets the specifications given in this
instruction manual,
Note that an output voltage of maximum 5.5 V DC is supplied by a
device connected via USB.
Attention
Use of a multiple socket extension cord
Risk of personal injury or material damage caused by unsafe multiple
socket extension cord
If you use a multiple socket extension cord to connect the
Corvis® ST to the power supply, you must heed the following information:
Use an extension cord that complies with the requirements of DIN
EN ISO 60601-1: 20005, section 16.
Do not place the multiple socket extension cord on the floor.
Do not use more than one multiple socket extension cord.
Plug only the Corvis® ST and the computer that is being used with
the unit (if applicable) into the multiple socket extension cord.
If you are using a multiple socket extension chord it has to be supplied
with a isolation transformer.
If you are using a new computer for the Corvis® ST, you must have the
electrical safety checked. Call OCULUS Service for this purpose.
Electromagnetic Compatibility (EMC) / Cables
Risk of personal injury or damage to property due to electromagnetic in-
terference
Portable and mobile RF communications equipment can affect medical
electrical equipment
.
Make sure that portable and mobile RF communications equipment
do not cause interference.
Recommendation: Maintain a minimum distance of 4 m. If the
distance is shorter, you must ensure that the Corvis® ST functions
correctly.