4 Safety In
s
truction
s
Instruction Manual Myopia Master® (G/68100/EN Rev04 0820)
7 / 92
In the patient environment, use devices that conform to IEC 60601-1. If
a multiple socket outlet is to be used, or if a device is to be used that does
not meet the IEC 60601-1 standard, use an isolating transformer.
Information about the operation of an ME system
The Myopia Master® and a connected computer form a medical
electrical system (ME system) according to IEC 60601-1. If you connect
additional devices, such as, for example a printer, those devices become
part of the ME system.
Make sure that all devices of the ME system meet the requirements
of IEC 60601-1, IEC 60950-1 or 62368-1.
Note that an output voltage of maximum 5.5 V DC is supplied by a
device connected via USB.
Instructions for Operation
Never operate a damaged Myopia Master®.
Only operate the Myopia Master® with the original accessories
supplied by us and only when the unit is in technically perfect
condition.
Before first use: Let OCULUS or an authorized dealer train you in the
operation of the Myopia Master®.
Only operate the device if you have understood the operating
instructions.
Fig. 4-1: Patient environment