background image

 

 

25 

Technique – OLIV-WFNA, OLIV-EQNA 

 

The lens may be held on the eye by an assistant using the 
Landers Lens Handle (OLIV-H) or by suturing one of the 
Landers Lens Rings to the sclera. 

After a suitable wetting agent is placed on the cornea, the 
lens is placed on the cornea. 

Many surgeons do not use an inverted image contact lens until the anterior third of the vitreous has been 
removed and a deeper image of the vitreous cannot be obtained with normal microscope observation. 

Turn off the coaxial and oblique illumination  of the microscope, since this  may lead to  reflections  from 
the contact lens surfaces. Check the positions of instruments repeatedly before and during the operation, 
as it is very difficult to recognize the patient’s crystalline lens through a contact lens.  

In  order  to  focus  the  microscope,  set  it  to  its  lowest  magnification  and  then  raise  the  microscope  head 
away  from  the  patient’s  eye.  It  is  suggested  that  one  work  at  the  lowest  magnification.  Most  surgeons 
reduce magnification after they become familiar with the IVS so they may achieve more field of view. 

Be  sure  the  lens  is  seated  well  on  the  cornea.  If  the  assistant  has  a  poor  image  and  you  find  the  image 
good  (or  vice  versa),  it  is  possible  only  one  observation  beam  path  of  the  microscope  is  receiving  and 
transmitting a good image. Slightly shifting the lens will correct the problem 

You  can  bring  the  pars  plana  into  view  by  tilting  the  contact  lens  a  little  or  by  shifting  it  horizontally. 
Some  lenses  possess  large  depth  of  field.  The  concavity  of  the  fundus  then  appears  slightly  flattened, 
especially toward the periphery. It is also possible that the anterior parts of intraocular instruments will at 
first appear somewhat thicker and slight bent or curved. 

Keep endo-illumination as far as possible from the retina and increase illumination at its tip. This utilizes 
the wide-angle effect of the Wide Field Lens to its fullest. Light intensity at the retina will be somewhat 
reduced due to the distance from the retina. 

Fluid/gas  exchanges,  fluid/silicone  exchanges  and  gas/silicone  exchanges  can  be  easily  be  optically 
monitored even in phakic eye with the Wide Field Lens. 

Summary of Contents for IVS2

Page 1: ...OCULAR INSTRUMENTS IVS2 INVERTER VITRECTOMY SYSTEM PRODUCT MANUAL...

Page 2: ......

Page 3: ...al Tips on using the IVS2 Page 13 Maintenance Removable Accessories Care Cleaning and Disinfection of the IVS2 Sterilization of the Adjustment Knob Page 14 Troubleshooting Page 15 Guarantee Limitation...

Page 4: ...treatment of diseases and disorders of the eye The Inverter Vitrectomy System re inverts the inverted image of our line of Wide Angle Vitrectomy Lenses Proper installation and care of the device is i...

Page 5: ...cally flawless working condition Do not use the IVS2 if it is defective Please contact Ocular Instruments or its distributor if you believe the unit to be defective Do not use the IVS2 in the presence...

Page 6: ...quatorial II Vitrectomy Lens OLIV EQ 2 1 Woldoff High Magnification Vitrectomy Lens OWIV HM OR a combination of the following non autoclavable lenses 1 Landers Non Autoclavable Wide Field Vitrectomy L...

Page 7: ...ce the fundus is observed through a contact lens it is necessary to be able to move back and forth smoothly and quickly between the upright and the inverted image This function is performed by the IVS...

Page 8: ...have the knowledge and experience to ensure proper handling The IVS2 is intended for use with operating microscopes in hospitals clinics outpatient facilities etc where ophthalmic surgery is performe...

Page 9: ...er Adjustment Knob sterilizable Cap for adjustment knob No 3 above No 2 Knurled thumb screw Fastens upper part of microscope to the IVS2 No 5 Protective dust cover No 3 Adjustment knob No 6 Rating pla...

Page 10: ...S2 like the operating microscope should be handled with care and should not be subjected to shocks high temperature moisture or fluids or high temperatures above 40 C Before assembling the IVS2 and mi...

Page 11: ...e locking screw on the main body of the microscope Figure 2 No 1 with the screwdriver provided Remove the binocular eyepiece assembly and the beam splitter from the main body of the microscope On many...

Page 12: ...d other components of the microscope that is carefully tighten the knurled thumb screw of the IVS2 Fig 1 No 2 in the same manner and then confirm that each component is correctly seated Figure 3 Micro...

Page 13: ...11 Illustrations of IVS2 on Various Microscopes Zeiss Microscope...

Page 14: ...12 Leica Wild Microscope...

Page 15: ...13 Topcon Microscope...

Page 16: ...Knob OIVS2 K either clockwise or counterclockwise to its other position Always activate the IVS2 if you are using optical inversion e g with Ocular Instruments steam sterilizable or non autoclavable...

Page 17: ...m autoclavable at 270 F 132 C for the appropriate time period per your sterilizer s instructions Transport of the IVS2 Always position the optical system of the IVS2 in one of the two rest positions O...

Page 18: ...slightly Seeing surgeon s hand movements outside the image of the eye causes nausea The IVS2 inverts all of the images that come into the microscope Since the wide angle lenses invert the image of the...

Page 19: ...se return it to Ocular Instruments in it s carrying case with both protection caps covering the optics Proof of purchase and a copy of the original packing slip are required before services can be pro...

Page 20: ...es or repairs are made by unauthorized persons or if the unit is improperly serviced or handled Service Address Supplementary information is available from our service department or from our authorize...

Page 21: ...Pa Degree of waterproofing IP 64 The IVS2 is adaptable to the following microscopes Zeiss OPMI 1 6 Retrolux 1 3 CS Retroskop 1 2 CS MDI MDO MDU Leica M 650 M500 501 M 690 M620 M690 M 840 M841 M820 M84...

Page 22: ...deotaping important procedures Landers Equatorial II Vitrectomy Lens Designed with Maurice B Landers III M D Chapel Hill NC Product Code OLIV EQ 2 The Landers Equatorial Vitrectomy Lens is excellent f...

Page 23: ...nd then raise the microscope head away from the patient s eye It is suggested that one work at the lowest magnification Most surgeons reduce magnification after they become familiar with the IVS so th...

Page 24: ...through a contact lens With the microscope set at the lowest magnification and the microscope head travel at the lowest position move the microscope to obtain a focus on the cornea Once the lens has...

Page 25: ...ach half of the lens with mild soap and water so that each element is free of mucous sebaceous deposits or other debris Caution If fluid gas exchange has occurred wipe lens with alcohol to remove any...

Page 26: ...Maurice B Landers III M D Chapel Hill NC Product Code OLIV EQNA The Landers NA non autoclavable Equatorial Vitrectomy Lens is a single piece lens designed for clinical situations where autoclaving is...

Page 27: ...reduce magnification after they become familiar with the IVS so they may achieve more field of view Be sure the lens is seated well on the cornea If the assistant has a poor image and you find the im...

Page 28: ...cation and the microscope head travel at the lowest position move the microscope to obtain a focus on the cornea Once the lens has been placed on the eye focus the microscope using the focusing adjust...

Page 29: ...rsilene suture the suture needle is inserted immediately next to the limbus at the 3 00 and 9 00 position using a single armed suture The suture is then cut midway between the 3 00 and 9 00 posts The...

Page 30: ...28 Blank Page...

Page 31: ...29 Blank Page...

Reviews: