10
Now place the flange of the IVS2 into the dovetail receptacle of the main body
of the microscope. Check that the IVS2 is properly seated before carefully
tightening the locking screw on the main body of the microscope (Fig 3, No.
1). When properly mounted, the IVS2 should not wobble. Make sure of this
before attaching the other microscope components.
Proceed in the same manner when mounting the beam splitter, the binocular
eyepiece and other components of the microscope; that is, carefully tighten the
knurled thumb screw of the IVS2 (Fig 1, No. 2) in the same manner and then
confirm that each component is correctly seated.
Figure 3 - Microscope After Installation of IVS2
No. 1
Microscope Locking Screw
If an endolaser is used, the laser shutter can be positioned either between the
main body of the microscope and the IVS2 or between the IVS2 and the beam
splitter.
A separate eye protector for the assistant must be provided if the laser
shutter can only be positioned above the beam splitter.
When using a laser, always follow the instructions of the laser manufacturer.