background image

Gordon Smart Recorder User Guide 

Gordon Quick Start 

1.

 

Provide power to the Battery Pack  

a.

 

Insert batteries into Battery Pack (see Battery Pack User Manual) 

b.

 

Move toggle switch on the Gordon Board to the ON position (as seen below)  

 

2.

 

Attach hydrophone(s) to battery pack using a cable or directly in connector 

3.

 

Connect to the Gordon Board using a micro USB connector provided to set up for deployment 
(see 

Appendix A

) 

a.

 

Connect the micro USB cable to Gordon Board 

b.

 

Connect the opposite end of the cable to a USB port on your PC 

4.

 

Install the Gordon application provided on the USB drive (see 

Appendix B

5.

 

Open the Gordon application on your PC 

6.

 

Choose Connect to a Smart Recorder  

   

 

7.

 

Choose your device from the list: 

a.

 

Click on Battery Pack  

b.

 

Click Connect 

 

 

 

ON 

OFF 

Summary of Contents for Gordon

Page 1: ...Gordon Smart Recorder User Guide January 2019 Ocean Sonics Ltd Hill House 11 Lornevale Road Great Village NS B0M 1L0 Canada Phone 1 902 655 3000 www OceanSonics com...

Page 2: ...Batteries 8 Dry Run Deployment Test Using Batteries 9 Appendix A 10 Connecting micro USB to Gordon Board 10 Appendix B 11 Installing the Gordon Application 11 Appendix C 12 Actions Menu 12 Set the Sma...

Page 3: ...with a Gordon Board inside Gordon is used to manage the recording of data With the ability of additional memory from a hard drive and a low power duty cycling it allows longer amounts of time to be re...

Page 4: ...le or directly in connector 3 Connect to the Gordon Board using a micro USB connector provided to set up for deployment see Appendix A a Connect the micro USB cable to Gordon Board b Connect the oppos...

Page 5: ...he date presumed by the Smart Recorder must be set below If there is synchronization from the GPS the time source that the Smart Recorder is using to set the time the model and the firmware version of...

Page 6: ...ensure the proper number of channels have been added for the number of hydrophones on the battery pack e Choose the settings for the Waveform Data Spectrum Data Delay and Smart Recorder Settings f A S...

Page 7: ...he battery power and memory capability of your system Use the excel sheet provided on the USB with the Gordon Smart Recorder to help choose the settings that will provide the most useful data from you...

Page 8: ...tart up of the hydrophones is indicated by one buzz followed by a double buzz The shut down of hydrophones will be indicated by three consecutive buzzes The minimum period suggested is 10 minutes Any...

Page 9: ...en provided Plug into the top board as shown in Step 5 1 Remove any batteries from the Battery Pack 2 Set up the Battery Pack in clean dry environment so top board Gordon Board is accessible 3 Lay the...

Page 10: ...uick Start instructions above 2 Allow the hydrophone to record at least two or more duty cycles 3 Verify data was recorded properly on chosen memory option 4 After testing ensure all instruments are t...

Page 11: ...mart Recorder User Guide Appendix A Connecting micro USB to Gordon Board 1 Connect the micro USB to the port beside the toggle switch on the Gordon Board 2 Connect the other end into a USB port on you...

Page 12: ...d Paste the Gordon_v1_0_2_Setup exe onto PC Desktop or into a chosen folder 2 Double Click on the copied file to run the Installation Setup 3 Click through the following Setup procedures a Click Next...

Page 13: ...and Add Channel IP Addresses The Smart Recorder sets up a local Network from the smart recorder to the hydrophone s The IP addresses for the hydrophones will automatically be set by the Smart Recorder...

Reviews: