MNL-1017 Rev A
6
Verify Calibration
•
Press
“
Sample to Water
”.
•
Press
“
Measure
”
•
Verify that the temperature did
not change more than 0.2°C.
o
Calibration and Current
temps should be within
0.2°C for each
spectrometer (S0 and Ref).
See arrows.
•
Make sure transmission
measurement is between
99.5% and 100.5% at all
wavelengths
•
If not, recalibrate and repeat.