SPECIFICATIONS

Guidance and manufacturer’s declaration - electromagnetic immunity

The bed is intended for use in the electromagnetic environment specified below. The customer or the 

user of the bed should assure that it is used in such an environment.

Immunity Test

IEC 60601 test level

Compliance level

Electromagnetic

environment - guidance

Electrostatic discharge 

(ESD)

IEC 61000-4-2

±6 kV contact

±8 kV air

±6 kV contact

±8 kV air

Floors should be wood, 

concrete or ceramic tile. 

If floors are covered with 

synthetic material, the rela-

tive humidity should be at 

least 30 %.

Electrical fast transient/

burst

IEC 61000-4-4

±2 kV for power supply lines

±1 kV for input/output lines

±2 kV for power supply lines

±1 kV for input/output lines

Mains power quality should 

be that of a typical com-

mercial or hospital environ-

ment.

Surge

IEC 61000-4-5

±1 kV differential mode

±2 kV common mode

±1 kV differential mode

±2 kV common mode

Mains power quality should 

be that of a typical com-

mercial or hospital environ-

ment.

Voltage dips, short inter-

ruptions and voltage 

variations on power supply 

input lines IEC 61000-4-11

<5 % UT

(>95 % dip in UT)

for 0,5 cycle

40 % UT

(60 % dip in UT)

for 5 cycles

70 % UT

(30 % dip in UT)

for 25 cycles
<5 % UT

(>95 % dip in UT)

for 5 sec

<5 % UT

(>95 % dip in UT)

for 0,5 cycle

40 % UT

(60 % dip in UT)

for 5 cycles

70 % UT

(30 % dip in UT)

for 25 cycles

<5 % UT

(>95 % dip in UT)

for 5 sec

Mains power quality should 

be that of a typical com-

mercial or hospital environ-

ment. If the user of the bed 

requires continued opera-

tion during power mains 

interruptions, it is recom-

mended that the bed be 

powered from an uninter-

ruptible power supply or a 

battery.

Power frequency (50/60 

Hz) magnetic field IEC 

61000-4-8

3 A / m

3 A / m

Power frequency magnetic 

fields should be at levels 

characteristic of a typical 

location in a typical com-

mercial or hospital environ-

ment.

NOTE U

T

 is the a.c. mains voltage prior to application of the test level.

Summary of Contents for PX150

Page 1: ...USER MANUAL www oakworksmed com 717 235 6807 made in the USA with US imported parts PX150 PX200 EMERGENCY FIELD BEDS...

Page 2: ...within this document is subject to change without notice and should not be construed as a commitment by OAKWORKS Inc OAKWORKS Inc encourages requests for technical specifica tions and the like documen...

Page 3: ...Back Rest 8 Side End Rail Use 8 Moving the bed 9 Dual Locking Casters Use 9 Cleaning Disinfection Recommended Cleaners Disinfectants 10 Cleaning Process 11 Inspections Maintenance 11 Warranty Informat...

Page 4: ...al or other medical facility The bed is only to be moved within the patient room for cleaning or patient access No spe cial training is required but a review of the following manual is important for t...

Page 5: ...injury to persons 1 Unplug this furnishing from the electrical outlet before cleaning 2 Unplug from outlet before adding or removing parts 3 Never operate this furnishing if it has a damaged cord or...

Page 6: ...Otherwise individuals may become trapped between the bed rail and the bed frame Bed rails can be deformed or broken if excessive side pressure is exerted on the bed rails These bed rails are used for...

Page 7: ...st Mattress 6 Thick 15 cm Removable Water Resistant Foot Control Height Motion Bed Voltage Options 120V 60Hz or 230V 50Hz Bed Weight 180 lbs 82 kg Bed Capacity 400 lbs 182 kg patient and accessories W...

Page 8: ...rest Width 36 91 cm Bed Weight 200 lbs 91 kg 4 Button Hand Control Height Backrest Padding 6 Thick 15 cm Removable Water Resistant Bed Voltage Options 120V 60Hz or 230V 50Hz Bed Capacity 400 lbs 182 k...

Page 9: ...plugged into an appropriate outlet that is prop erly installed and grounded in accordance with all local codes and ordinances See U S sample below Correct Implementation Incorrect Implementation Risk...

Page 10: ...CONTROL OPERATION DIRECTIONS FOR USE Height DOWN Height UP The Oakworks PX200 bed offers height and backrest positioning with electronic ease Operate the Hand Control as shown below to raise or lower...

Page 11: ...e of their weight off the back rest while you lift CAUTION ADJUSTING THE BACK REST SECTION PX150 BEDS ONLY Do not sit on the Backrest section 8 DIRECTIONS FOR USE SIDE END RAIL USE Firmly grab the han...

Page 12: ...to protect the surface from any debris from the bottom of your foot control Two people are required to safely move the bed Grasp the bed frame securely at the end with either the locking casters or t...

Page 13: ...KWORKS recommends a prepackaged wipe for cleaners disinfectants to ensure best distribution of disinfectant for the required kill time without leaving excess residue and or overexposing components the...

Page 14: ...he bed 2 Throw the used wipe s away 3 Using a fresh wipe or wipes thoroughly wipe all surfaces of the top and any high contact areas such as handles handsets etc making sure they remain wet for the di...

Page 15: ...lways start with a letter The model number and serial number are located on the underside of the seat section Model Number Serial Number 16796025 PXEXFT311873ACTTVN Ser PXR563146 PX FT 30 x 22 38 x 72...

Page 16: ...amp 50 Hz Current Draw 2 7 amps 1 5 amps 3 3 amps Maximum Momentary Current Consumption 9 0 amps 5 0 amps 11 0 amps Voltage Output to Actuators 24 VDC 24 VDC 24 VDC Electric Shock Protection Class 1...

Page 17: ...c environment in which radiated RF disturbances are controlled the customer or the user of the bed can help prevent electromagnetic interference by maintaining a minimum distance between portable and...

Page 18: ...al mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical com mercial or hospital environ ment Voltage dips short inter ruptions and voltage vari...

Page 19: ...RF transmitters as deter mined by an electromagnetic site survey ashould be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the...

Page 20: ...reedom PA 17349 Phone 717 235 6807 FAX 717 235 6798 www oakworksmed com made in the USA with US imported parts 3034177 CONFORMS TO ANSI AAMI STD ES60601 1 IEC STD 60601 1 3RD EDITION IEC STD 60601 1 2...

Reviews: