34
SPECIFICATIONS
PRODUCT SPECIFICATIONS
NORTH AMERICA
EUROPE
Weight
359-452 lbs.
163-205 kg.
Shipping Weight
452-544 lbs.
205-247 kg.
Lifting Capacity
500 lbs. (750 lbs. Bariatric)
227 kg. (340 kg. Bariatric)
Materials of Construction
This product contains no latex.
Aluminum Filtration
Equivalence and thickness
of Carbon Fiber Tabletop
ALUMINUM EQUIVALENCE
THICKNESS
Non-Bariatric Tops
100 kVp, HVL of 3.6mm = 1.10 (1.35 max)
1.65” (4.2 cm)
Bariatric Tops
100 kVp, HVL of 3.6mm = 1.50 (1.79 max)
2.18” (5.5 cm)
All Extensions
100 kVp, HVL of 3.6mm = 1.66 (2.00 max)
.5” (1.3 cm)
1” (2.5 cm.) Table Pad
100 kVp, HVL of 3.6mm = .88 (1.07 max)
1” (2.5 cm)
2” (5 cm.) Table Pad
100 kVp, HVL of 3.6mm = .88 (1.07 max)
2” (5 cm)
This product complies with United States Department of Health and Human Services radiation performance standards,
21 CFR Sub chapter J, in effect at the time of manufacture for radiographic tables.
ENVIRONMENTAL CONDITIONS
Conditions
Temperature
Humidity
Atmospheric Pressure
Normal Use
50° (10°C) to 104° (40°C)
20% to 60% RH
98 to 105 kPa
Storage & Transport
-20° (-29°C) to 135° (57°C)
20% to 95% RH
98 to 105 kPa
ELECTRICAL SPECIFICATIONS
CFPM100 Tables Only
CFPM400, CFPM401, CFPM301,
CFLU401, CFUR401 & CFUR301
Tables
Designed for:
NORTH AMER-
ICA
EUROPE
NORTH AMER-
ICA
EUROPE
Input Service (All electric
tables)
120 VAC/15 amp/60
Hz
230 VAC/10
amp/50/60 Hz
120 VAC/15 amp/60
Hz
220 VAC/10
amp/50/60 Hz
Current Draw
5.8 amps
3.0 amps
8.2 amps
4.5 amps
Maximum Momentary Cur-
rent Consumption
9.0 amps
4.5 amps
12.3 amps
6.7 amps
Voltage Output to Actuators
120 VDC
220 VDC
24V top,
120V Column
24V top,
220V Column
Electric Shock Protection
Class 1 Equipment
Tabletop Applied Part
Type B Applied Part
Ingress Protection Rating
IPX0
Mode of Operation
Intermittent Operation: MAX 2 minutes ON / 18 minutes off