Chapter 2
OaktonLog Plus
152
2.5.5.
Sensor Calibration
A new OaktonLog comes fully calibrated. After a period of long
use, however, you may want to re-calibrate the humidity or the
temperature sensors.
The humidity calibration affects both OaktonLog and OaktonLog
plus and should be carried out while OaktonLog is connected to the
PC. The temperature calibration affects only the OaktonLog Plus
software and can be carried out regardless of whether OaktonLog
is connected to the PC.
1.
Calibration password
To prevent accidental change of the calibration, the calibration
procedure is protected by a password. The default password is:
1234. To change the password:
Summary of Contents for RH/TempLog
Page 1: ...TempLog RH TempLog User Guide Seventh Edition First Print Printed in July 2006...
Page 2: ......
Page 15: ...Software 205 Hardware 209 Appendix A Figures 210 Index 211...
Page 16: ......
Page 19: ...OaktonLog Chapter 1 19 Chapter 1 OaktonLog...
Page 27: ...OaktonLog Chapter 1 27 Run Cyclic Run Timer Run Stop Mode...
Page 97: ...OaktonLog Plus Chapter 2 97 Chapter 2 OaktonLog Plus...
Page 155: ...OaktonLog Plus Chapter 2 155 6 Click Default to restore the original values...
Page 166: ...Chapter 3 The Cradle 166 Chapter 3 The Cradle...
Page 187: ...External Magnet Antenna Chapter 5 187 Antenna jumper...