Oakton ION 5 Acorn Series Operating Instructions Manual Download Page 7

NOTE

: This meter has automatic buffer recogni-

tion which identifies the correct pH buffer values
during calibration. If buffers other than 4.01, 7.00
or 10.00 are used, or the electrode has worn out,
the LCD will flash "Er1".

See figure

Press CAL key to clear.

Temperature calibration 

(ION 6 meter only)

The temperature sensor included with your meter is factory calibrated. Over time,
the temperature calibration may drift and require recalibration. If you replace the
probe you should calibrate temperature prior to pH calibration.

1. 

Connect your temperature probe to the meter.

2. 

Press the MODE key until "°C" appears 
in the LCD.

3. 

Compare displayed value to a NIST certified
thermometer or other thermometer known 
to be accurate. For best accuracy, place 
probe and thermometer in a constant 
temperature bath.

4. 

Press the CAL key. The LCD shows "CA"
and the reading flashes. 

See figures

5. 

Use the 

and 

keys to adjust the 

temperature shown to match the accurate
temperature (maximum adjustment is 
±5°C from factory default).

6. 

Press ENTER key to confirm calibration.
The LCD displays "CO", and the meter 
then reverts to measurement mode.

13

12

4.4

4.5

C

A

 C

B

C

A

B

Resetting calibrated values to default values

When you calibrate the pH mode of the Ion 6 meter, old pH calibration data is 
written over on a point by point basis. If when recalibrating you do not calibrate at
all three points, the old data may result in errors. You can reset the meter to factory
calibration values using the following procedure. 

1. 

TURN METER OFF. 

Meter must be OFF in order to reset calibration values.

2. 

Press and hold the CAL key while switching on the meter using the ON key.
Hold the CAL key until the display shows "rSt". 

3. 

Press ENTER key to confirm. The meter automatically clears all stored pH 
calibration values and reverts to Ion measurement mode. 

Press MODE key instead of ENTER if you do not wish to reset. Meter will maintain
current calibration values and return to the Ion measurement mode.

Summary of Contents for ION 5 Acorn Series

Page 1: ...ON alone will determine if the product problem is due to deviations or customer misuse Out of warranty products will be repaired on a charge basis 11 Return of items Authorization must be obtained fro...

Page 2: ...on 12 4 5 Resetting the meter to factory calibration values 13 5 Measurement 14 15 5 1 Measuring ion concentration 14 5 2 Measuring mV 14 5 3 Measuring pH and temperature 15 5 4 Measuring pH with manu...

Page 3: ...ies Warranty and Return of Items section 2 Display and Keypad Functions The Acorn has a large custom LCD with the following indicators pH HO LO C mV indicates operation mode pH annunciator ION 6 only...

Page 4: ...tor push together and twist the electrode connector 1 2 turn until it clicks into place To remove simply rotate the connector counter clockwise until it unlocks and slide the connector off the socket...

Page 5: ...libration This instrument is capable of up to three point ion calibration Calibration values are 0 1 1 0 10 0 or 100 0 You must calibrate to at least two points All calibration points should be one de...

Page 6: ...s in the solution can affect the calibration and the accuracy of the measurements Before use remove the electrode soaker bottle If the electrode has been stored dry condition the glass bulb by soaking...

Page 7: ...he and keys to adjust the temperature shown to match the accurate temperature maximum adjustment is 5 C from factory default 6 Press ENTER key to confirm calibration The LCD displays CO and the meter...

Page 8: ...in pH and temperature 1 Rinse both electrode and temperature sensor with distilled water 2 Turn meter on ION 6 meter automatically enters ion measurement mode Press MODE key until meter enters pH meas...

Page 9: ...ed solution for about 10 minutes Allow the electrode to cool in some unheated KCl solution Protein deposits prepare a 1 pepsin solution in 0 1M of HCl Set the pH electrode in the solution for five to...

Page 10: ...ps and dings Also features meter stand for convenient tabletop use Shpg wt 0 5 lb 230 g Ion selective electrodes and solution kits WD 35802 00 Ammonia ion selective electrode includes 15 ml of referen...

Page 11: ...alibration solution pouches 20 box 0 01 pH accuracy at 25 C Shpg wt 1 1 lb 454 g per box WD 35653 00 Deionized rinse water solution pouches WD 35653 01 pH 4 01 Singles buffer solution pouches WD 35653...

Page 12: ...23 22...

Reviews: