Oakton ION 5 Acorn Series Operating Instructions Manual Download Page 6

11

10

pH buffer set selection (ION 6 meter only)

The ION 6 meter is capable of up to three-point pH calibration to ensure accuracy
across the entire range of the meter. The ION 6 meter can be set to recognize either
of two buffer sets: USA or NIST buffers.

USA

—“USA”: 4.01; 7.00 and 10.01

NIST

–“nSt”: 4.01, 6.86, and 9.18

To choose your buffer set:

1.

Turn meter OFF. Meter must be OFF in order to
access buffer selection.

2.

While holding down the MODE key, press
the ON/OFF key momentarily. Continue to
hold the MODE key until the display flashes
“buF”.

See figure

3.

Press the ENTER key to enter buffer set
selection mode. The display will flash
“USA”

See figure

4.

Press the MODE key to toggle between the
available buffer sets. The display will flash
“USA” or “nSt” as you toggle between 
the sets

5.

Press ENTER key to confirm the buffer set
you wish to use. Proceed with calibration 
in section 4.3

pH calibration (ION 6 meter only)

We recommend you perform at least a 2-point calibration at room temperature using
standard buffers that bracket (one above and one below) the expected sample range.
You can also perform a 1-point calibration, but make sure that the buffer value is
close to the sample value you are measuring.

All new calibrations will over-ride existing data. 

NOTE

Do not reuse buffer solutions after calibration. Contaminants in the 

solution can affect the calibration, and the accuracy of the measurements.

Before use: 

remove the electrode soaker bottle. If the electrode has been stored dry,

condition the glass bulb by soaking it in tap water for 30 minutes. This hydrates the
glass bulb if the electrode is too dry, or has not been used for a long period of time.

1.

Turn meter on. Meter will automatically enter ion measurement mode. Press the
MODE key until display shows “pH” in the upper right corner.

2.

Rinse electrode thoroughly with deionized water or a rinse solution. 
DO NOT wipe the electrode;

this causes a build-up of electrostatic 

charge on the glass surface.

3. 

Dip both the electrode and temperature sensor into pH 7.00 (or 6.86 in NIST set)
buffer solution. The glass bulb must be completely immersed into the sample.
Stir gently, and wait for reading to stabilize (approximately 40 seconds). 

4. 

Press CAL key to enter the calibration mode. 
The display will momentarily flash “CA” to indicate CAlibration.

See figure

The display will then show the current 
noncalibrated reading, blinking while in 
calibration mode.

See figure

5. 

Allow the reading to stabilize. The meter 
automatically recognizes 7.00, 4.01 or 10.00
buffers (or 4.01; 6.86 and 9.18 in NIST set).

6. 

Press ENTER key once to confirm calibration.
The LCD displays “CO” to indicate the 
calibration point has been confirmed. The
meter exits calibration mode and returns 
to measurement mode.

7. 

Repeat with pH buffers 4.01 and/or 10.00 
( or 6.86/9.18) for best accuracy.

4.2

A

B

pH

A

B

4.3

A

B

A

B

Summary of Contents for ION 5 Acorn Series

Page 1: ...ON alone will determine if the product problem is due to deviations or customer misuse Out of warranty products will be repaired on a charge basis 11 Return of items Authorization must be obtained fro...

Page 2: ...on 12 4 5 Resetting the meter to factory calibration values 13 5 Measurement 14 15 5 1 Measuring ion concentration 14 5 2 Measuring mV 14 5 3 Measuring pH and temperature 15 5 4 Measuring pH with manu...

Page 3: ...ies Warranty and Return of Items section 2 Display and Keypad Functions The Acorn has a large custom LCD with the following indicators pH HO LO C mV indicates operation mode pH annunciator ION 6 only...

Page 4: ...tor push together and twist the electrode connector 1 2 turn until it clicks into place To remove simply rotate the connector counter clockwise until it unlocks and slide the connector off the socket...

Page 5: ...libration This instrument is capable of up to three point ion calibration Calibration values are 0 1 1 0 10 0 or 100 0 You must calibrate to at least two points All calibration points should be one de...

Page 6: ...s in the solution can affect the calibration and the accuracy of the measurements Before use remove the electrode soaker bottle If the electrode has been stored dry condition the glass bulb by soaking...

Page 7: ...he and keys to adjust the temperature shown to match the accurate temperature maximum adjustment is 5 C from factory default 6 Press ENTER key to confirm calibration The LCD displays CO and the meter...

Page 8: ...in pH and temperature 1 Rinse both electrode and temperature sensor with distilled water 2 Turn meter on ION 6 meter automatically enters ion measurement mode Press MODE key until meter enters pH meas...

Page 9: ...ed solution for about 10 minutes Allow the electrode to cool in some unheated KCl solution Protein deposits prepare a 1 pepsin solution in 0 1M of HCl Set the pH electrode in the solution for five to...

Page 10: ...ps and dings Also features meter stand for convenient tabletop use Shpg wt 0 5 lb 230 g Ion selective electrodes and solution kits WD 35802 00 Ammonia ion selective electrode includes 15 ml of referen...

Page 11: ...alibration solution pouches 20 box 0 01 pH accuracy at 25 C Shpg wt 1 1 lb 454 g per box WD 35653 00 Deionized rinse water solution pouches WD 35653 01 pH 4 01 Singles buffer solution pouches WD 35653...

Page 12: ...23 22...

Reviews: