Oakton ION 5 Acorn Series Operating Instructions Manual Download Page 5

6.

If you DO wish to calibrate to this point, press the ENTER key. The meter
will display the mV output of your electrode in the standard solution. Allow
the instrument to stabilize. Stabilization may take a few minutes depending
on electrode and standard concentration. Electrode response in low concen-
trations is much slower than response in high concentrations.

7.

Press ENTER to confirm your calibration. The meter will move on to the 
next standard (1.0 ppm).

8. 

Follow steps 5-7 for each of the next calibration points. Make sure to rinse off 
the electrode with deionized water before placing it in the next calibration 
standard.

9. 

After you have entered three calibration
points, the meter will display the average
electrode slope in mV. 

See figure

You may compare this reading with the
expected slope value for your electrode from
your electrode manual to verify electrode
operation.

If you are only calibrating to two points, 
you will need to press the 

(up arrow) key

until the slope is displayed.

If you want to leave the calibration mode 
for any reason, press the CAL key to exit 
the calibration mode. The meter will revert
back to the earlier calibration curve.

You must calibrate to at least two points in 
order for calibration to take effect. If you only confirm the one calibration point, the
meter will display “Er2” when you exit calibration mode, and the meter will revert
back to the earlier calibration curve.

See figure

9

8

4. Calibration

Ion calibration

This instrument is capable of up to three-point ion calibration. Calibration values are
0.1, 1.0, 10.0 or 100.0. You must calibrate to at least two points. All calibration points
should be one decade apart. For example you may perform three point calibration to 
0.1, 1.0 and 10.0, or two point calibration to 10.0 and 100.0. Do NOT perform two-
point calibration to 1.0 and 100.0 (meter will display “Er4”). For best accuracy, 
calibrate your meter to points with similar concentrations to the solutions you 
want to test.

Before you begin calibration, you must know whether the ion you are measuring is
mono or divalent. If you are uncertain, consult your ISE electrode manual, or your
OAKTON distributor.

1.

Prepare at least two calibration standards in accordance with the electrode 
instructions or your approved methods. For best
results, make sure standards and samples are at
the same temperature

2.

Turn meter on. The ION 5 meter will automati-
cally enter the ion measurement mode.

See figure

3.

Dip electrode into the first calibration standard.
Make sure to start with the calibration standard
that has the lowest concentration and move up
to the standards that have higher concentrations.

4.

Press CAL key to enter calibration mode. The
display will show CA1. 

See figure

If you are measuring monovalent ions, press
ENTER to begin calibration.

If you are measuring divalent ions, press 
MODE. The display will show CA2. 
Press ENTER to begin calibration

See figure

5. 

The first calibration point to appear is 0.1 ppm.
If you do NOT wish to calibrate to this point,
press 

(up arrow) to skip, and continue on to

the next calibration point.

4.1

A

A

B

B

mV

D

D

E

E

C

C

Summary of Contents for ION 5 Acorn Series

Page 1: ...ON alone will determine if the product problem is due to deviations or customer misuse Out of warranty products will be repaired on a charge basis 11 Return of items Authorization must be obtained fro...

Page 2: ...on 12 4 5 Resetting the meter to factory calibration values 13 5 Measurement 14 15 5 1 Measuring ion concentration 14 5 2 Measuring mV 14 5 3 Measuring pH and temperature 15 5 4 Measuring pH with manu...

Page 3: ...ies Warranty and Return of Items section 2 Display and Keypad Functions The Acorn has a large custom LCD with the following indicators pH HO LO C mV indicates operation mode pH annunciator ION 6 only...

Page 4: ...tor push together and twist the electrode connector 1 2 turn until it clicks into place To remove simply rotate the connector counter clockwise until it unlocks and slide the connector off the socket...

Page 5: ...libration This instrument is capable of up to three point ion calibration Calibration values are 0 1 1 0 10 0 or 100 0 You must calibrate to at least two points All calibration points should be one de...

Page 6: ...s in the solution can affect the calibration and the accuracy of the measurements Before use remove the electrode soaker bottle If the electrode has been stored dry condition the glass bulb by soaking...

Page 7: ...he and keys to adjust the temperature shown to match the accurate temperature maximum adjustment is 5 C from factory default 6 Press ENTER key to confirm calibration The LCD displays CO and the meter...

Page 8: ...in pH and temperature 1 Rinse both electrode and temperature sensor with distilled water 2 Turn meter on ION 6 meter automatically enters ion measurement mode Press MODE key until meter enters pH meas...

Page 9: ...ed solution for about 10 minutes Allow the electrode to cool in some unheated KCl solution Protein deposits prepare a 1 pepsin solution in 0 1M of HCl Set the pH electrode in the solution for five to...

Page 10: ...ps and dings Also features meter stand for convenient tabletop use Shpg wt 0 5 lb 230 g Ion selective electrodes and solution kits WD 35802 00 Ammonia ion selective electrode includes 15 ml of referen...

Page 11: ...alibration solution pouches 20 box 0 01 pH accuracy at 25 C Shpg wt 1 1 lb 454 g per box WD 35653 00 Deionized rinse water solution pouches WD 35653 01 pH 4 01 Singles buffer solution pouches WD 35653...

Page 12: ...23 22...

Reviews: