Troubleshooting
Code
Problem
Action
– – – (on display)
Target temperature
Select target within specifications
is over or under range
Battery icon
Low battery
Check and/or replace battery
Blank display
Possible dead battery
Check and/or replace battery
Laser doesn’t work
(1) Low or dead battery
(1) Replace battery
(2) Ambient temperature
(2) Use in area with lower ambient
above 40°C (104°F)
temperature
ERR
Possible damage
Contact your distributor
by EMF
Maintenance
Lens Cleaning: Blow off loose particles using clean compressed air. Gently brush
remaining debris away with a camel's hair brush. Carefully wipe the surface with a
moist cotton swab. The swab may be moistened with water. NOTE: DO NOT use
solvents to clean the plastic lens.
Cleaning the housing: Use soap and water on a damp sponge or soft cloth. NOTE: DO
NOT submerge the unit in water.
CE Certification
This instrument conforms to the following standards:
• EN50081-1:1992, Electromagnetic Emissions
• EN50082-1:1997, Electromagnetic Susceptibility
Tests were conducted using a frequency range of 80-1000 MHz with the instrument
in three orientations. The average error for the three orientations is ±0.82°C
(±1.48°F) at 3 v/m throughout the spectrum. However, between 234 MHz and 684
MHz at 3 V/m, the instrument may not meet its stated accuracy.
English
10