14
SECTION 6:
TROUBLE SHOOTING
PROBLEM PROBABLE
CAUSE
REMEDY
Automatic shutter does not
open when light source is in
exposure position.
Trigger magnet not set
properly: electrical
connections; defective
electrical components.
Check electric circuit; reed
switch; adjust trigger magnet
position.
Vacuum chuck does not
float when chuck release is
on.
Insufficient air going to
vacuum/air bearing cup.
Adjust airflow using "chuck
level release" valve.
Vacuum chuck vibrates
when the chuck release is
on.
Excessive air going to
vacuum/air bearing cup.
Adjust airflow using "chuck
level release" valve.
Short lamp life; lamp
explosion
Improper cooling (too much
or too little).
Insure that airflow maintains
lamp at proper operating
temp.
NOTE
: A new lamp should operate in the idle mode at the voltage/current indicated
on its package. The proper operating temperature can be found by installing a new
lamp and adjusting the airflow until the listed voltage/current is indicated by the meter
on the front of the Constant Intensity Controller.