Initializing DNA Technology

®

 

1.

 

Power the Freedom on and allow it to run for five minutes. 
 

2.

 

From the 

Main Screen

, press the Menu (M) button. 

 

3.

 

Press the Plus (+) button to scroll to the        Cell Info icon and press the Menu (M) button to select. 
 

4.

 

To ensure the Cell connection is working properly, check the Cell status indicated by the red box in 
the image below. A well-

functioning Cell will display “Cell: Connected.” It may take a few minu

tes 

for the Cell to connect. 
 

5.

 

If the Cell status is other than “Cell: Connected,” check the Cell signal strength in

dicated by the 

green box in the image below. A signal strength below 15 has a reduced chance of a proper 

connection. You may need to relocate the Freedom within the building to improve signal strength.  
 

6.

 

To ensure that the GPS is working properly, check the geographic coordinates near the bottom of 
the screen indicated by the yellow box in the image below. The GPS will display the current 
latitude and longitude of the unit. These coordinates can be verified using a web mapping 
application such as Google Maps. 
 

7.

 

If the GPS does not display the F

reedom’s

 

correct location, confirm the unit has a “line of sight” to 

the sky. This may require you to move the unit near a large window or outdoors temporarily. It may 
take several minutes for the GPS to capture a position lock. 
 

8.

 

To return to the Main Screen, press the Menu (M) button; press the Plus (+) button to scroll to the          
Home Exit icon and press the Menu (M) button to select. 
 

9.

 

If the Freedom has not connected within 15 minutes, power off and remove all power sources 
(batteries, AC/DC power supply). After 30 seconds, reconnect the power supply and power the 
device on to repeat the process. 
 
 

 

                            

 

Device Monitoring 

1.

 

Access O2 Concepts

 DNA portal at

www.o2dnaportal.com

 

2.

 

Enter user name and password. 
 
 
 
 
 
 

Summary of Contents for Oxlife Freedom

Page 1: ...OxlifeFreedom ProviderOrientation Guide O2 CONCEPTS COM 877 867 4008 800 1063 REV A...

Page 2: ...e Identification 1 Control Panel 2 LCD Display 3 Oxygen Outlet Cannula Port 4 Exhaust Port 5 Rechargeable Battery 6 External Power Port 7 Air Inlet Ports Air Inlet Filters Understanding the Control Pa...

Page 3: ...power cord to the AC power converter brick Connect the power supply to a wall outlet first and then to the power connection port of the Freedom The external power source icon will be displayed on the...

Page 4: ...can be accessed using the device serial number on the O2 Concepts DNA portal at www o2dnaportal com Air Travel The Freedom is an approved portable oxygen concentrator as indicated by the requisite FAA...

Page 5: ...properly check the geographic coordinates near the bottom of the screen indicated by the yellow box in the image below The GPS will display the current latitude and longitude of the unit These coordi...

Page 6: ...y maintenance on your Freedom other than routine maintenance described below Any such attempt will void the warranty Air Inlet Filter Cleaning Routine weekly cleaning of the air inlet filter and perio...

Page 7: ...cted above Per manufacturer s instructions disinfect exterior and battery using either Briallianize Plastic Cleaner with SofKloth Oxycide Ecolab or equivalent Cannula Specification The recommended max...

Page 8: ...y in flashing red Description No Breath After 15 minutes in Auto Pulse mode with no breath detected device will shut down Low Purity Oxygen level has dropped below 72 Low Battery Battery power is depl...

Page 9: ...O2 CONCEPTS COM 877 867 4008...

Reviews: